See
[:SOURce<hw>]:LFOutput<ch>:SHAPe:PULSe:WIDTh
<Width>
Sets the pulse width of the generated pulse.
Parameters:
<Width>
float
Range:
1E-6 to 100
Increment: 1E-8
*RST:
5E-4
Example:
See
Example "Configuring the LF generator"
Manual operation:
See
[:SOURce<hw>]:LFOutput<ch>:SHAPe:TRAPeze:FALL
<Fall>
Selects the fall time for the trapezoid shape of the LF generator.
Parameters:
<Fall>
float
Range:
1E-6 to 100
Increment: 10E-9
*RST:
250E-6
Example:
See
Example "Configuring the LF generator"
Manual operation:
See
[:SOURce<hw>]:LFOutput<ch>:SHAPe:TRAPeze:HIGH
<High>
Sets the high time for the trapezoid signal of the LF generator.
Parameters:
<High>
float
Range:
1E-6 to 100
Increment: 10E-9
*RST:
250E-6
Example:
See
Example "Configuring the LF generator"
Manual operation:
See
[:SOURce<hw>]:LFOutput<ch>:SHAPe:TRAPeze:PERiod
<Period>
Sets the period of the generated trapezoid shape. The period determines the repetition
frequency of the internal signal.