Remote control commands
R&S
®
SMM100A
994
User Manual 1179.1341.02 ─ 05
:TEST<hw>:DIRect:BLOCk:WRITe
<BoardName>, <SubAddress>, <StartAddress>,
<DataBlock>
Direct mode for programming the blocks of the hardware assemblies directly.
The respective hardware assembly responds directly to the command; any safety
mechanisms are bypassed. This function is only available via remote control.
Setting parameters:
<BoardName>
string
<Board>, <SubAddr>, <Addr>, <Data>
<SubAddress>
integer
<StartAddress>
integer
<DataBlock>
string
Example:
TEST:DIR:BLOC:WRIT 'TEST_SCPI',0,0,#1812345678
Usage:
Setting only
:TEST<hw>:ALL:STARt
Usage:
Event
Starts the selftest. Use the command
to query the result.
:TEST<hw>:ALL:RESult?
Queries the result of the performed selftest. Start the selftest with
.
Return values:
<Result>
0 | 1 | RUNning | STOPped
*RST:
STOPped
Usage:
Query only
13.20
TRIGger subsystem
The
TRIGger
system contains the commands for selecting the trigger source for the
RF and LF sweep.
You can work with an internal or with an externally applied trigger signal. In this case,
use the commands in the
SOURce:INPut
subsystem to configure the signal.
The trigger system of the R&S
SMM100A is a simplified implementation of the SCPI
trigger system. The
TRIGger
system differs from the SCPI system as follows:
●
No
INITiate
command; the instrument behaves as if
INITiate:CONTinuous
ON
were set.
TRIGger subsystem