Maintenance, storage and disposal
R&S
®
SMM100A
1028
User Manual 1179.1341.02 ─ 05
IQ Source
Selects the test signal source.
"Sine"
Generates a sine waveform with frequency as set with the parameter
"Constant I/Q"
Uses a constant I/Q test signal.
For more details, see
Chapter 4.6.3.4, "Constant IQ test signal"
"ARB"
Selects a signal generated by the ARB. Select the waveform file with
the parameter
.
Remote command:
Gain
Sets the gain for a sine or a constant I/Q test signal.
Remote command:
Frequency
If "IQ Source = Sine", sets the frequency of the test signal.
Remote command:
:TEST:BB:GENerator:FREQuency<ch>
Select ARB File
Provides access to the standard "File Select" function of the instrument. The provided
navigation possibilities in the dialog are self-explanatory.
See also, chapter "File and Data Management" in the R&S
SMM100A User Manual.
Remote command:
16.3.4.4
How to set up the R&S
SMM100A for BNC connections tests
1. This function is password-protected. Unlock the protection level 1 to access it.
2. Select "System Config > Setup > Maintenance > Selftest > Connections".
(signal input).
5. On the instrument, connect the selected "Source" connector with the "Destination"
connector.
The selection list in the dialog shows you where to find the associated connector
on the instrument.
6. Select "Execute BNC Connections Test".
Performing maintenance tasks