Remote control commands
R&S
®
SMM100A
869
User Manual 1179.1341.02 ─ 05
Parameters:
<IqRatio>
float
The setting value can be either in dB or %. An input in percent is
rounded to the closest valid value in dB.
Range:
-4 to 4
Increment: 1E-4
*RST:
0
Default unit: dB | PCT (setting command) / dB (result value)
Example:
SOURce:BB:IMPairment:RF1:IQRatio:MAGNitude 10 PCT
SOURce:BB:IMPairment:RF1:IQRatio:MAGNitude?
// 0.848
// the value is returned in dB
Manual operation:
See
[:SOURce<hw>]:IQ:IMPairment:LEAKage:I
<I>
[:SOURce<hw>]:IQ:IMPairment:LEAKage:Q
<Q>
[:SOURce]:BB:IMPairment:BBMM<ch>:LEAKage:I
<I>
[:SOURce]:BB:IMPairment:BBMM<ch>:LEAKage:Q
<Q>
[:SOURce]:BB:IMPairment:IQOutput<ch>:LEAKage:I
<I>
[:SOURce]:BB:IMPairment:IQOutput<ch>:LEAKage:Q
<Q>
[:SOURce]:BB:IMPairment:RF<ch>:LEAKage:I
<I>
[:SOURce]:BB:IMPairment:RF<ch>:LEAKage:Q
<Q>
Determines the leakage amplitude of the I or Q signal component of the corresponding
stream
Parameters:
<Q>
float
Range:
-10 to 10
Increment: 0.01
*RST:
0
Example:
SOURce:BB:IMPairment:RF1:LEAKage:Q 4 PCT
Sets the leakage for the Q-component to 4 percent.
Manual operation:
See
[:SOURce<hw>]:IQ:IMPairment:QUADrature[:ANGLe]
<Angle>
[:SOURce]:BB:IMPairment:BBMM<ch>:QUADrature[:ANGLe]
<Angle>
[:SOURce]:BB:IMPairment:FADer<ch>:QUADrature[:ANGLe]
<Angle>
[:SOURce]:BB:IMPairment:IQOutput<ch>:QUADrature[:ANGLe]
<Angle>
[:SOURce]:BB:IMPairment:RF<ch>:QUADrature[:ANGLe]
<Angle>
Sets a quadrature offset (phase angle) between the I and Q vectors deviating from the
ideal 90 degrees.
A positive quadrature offset results in a phase angle greater than 90 degrees.
SOURce subsystem