Remote control commands
R&S
®
SMM100A
970
User Manual 1179.1341.02 ─ 05
:SYSTem:RCL
<Pathname>
Loads a file with previously saved R&S
SMM100A settings.
Loads the selected file with previously saved R&S
SMM100A settings from the default
or the specified directory. Loaded are files with extension
*.savrcltxt
.
Setting parameters:
<Pathname>
string
Example:
SYSTem:RCL "/var/user/temp/Test"
Loads the file
Test.savrcltxt
from the direc-
tory
/var/user/temp/
.
Usage:
Setting only
:SYSTem:PROTect<ch>[:STATe]
<State>[, <Key>]
Activates and deactivates the specified protection level.
Suffix:
<ch>
.
Indicates the protection level.
See also
Parameters:
<State>
1 | ON | 0 | OFF
*RST:
n.a. (factory preset: 1)
Setting parameters:
<Key>
integer
The respective functions are disabled when the protection level
is activated. No password is required for activation of a level.
A password must be entered to deactivate the protection level.
The default password for the first level is 123456.
This protection level is required to unlock internal adjustments
for example.
Example:
To activate protection level:
SYSTem:PROTect1:STATe 1
Internal adjustments or hostname cannot be changed.
To unlock protection level 1:
SYSTem:PROTect1:STATe 0,123456
Internal adjustments are accessible.
Manual operation:
See
:SYSTem:SECurity:VOLMode[:STATe]
<SecPassWord>, <MmemProtState>
Activates volatile mode, so that no user data can be written to the internal memory per-
manently.
To enable volatile mode, reboot the instrument. Otherwise the change has no effect.