Remote control commands
R&S
®
SMM100A
698
User Manual 1179.1341.02 ─ 05
:OUTPut<hw>:AFIXed:RANGe:LOWer?
:OUTPut<hw>:AFIXed:RANGe:UPPer?
Queries the settable minimum/maximum value in mode
:OUTPut:AMODe FIXed
, i.e.
when the attenuator is not being adjusted.
Return values:
<Upper>
float
Increment: 0.01
Default unit: dBm
Example:
:OUTPut1:AMODe FIXed
:OUTPut1:AFIXed:RANGe:UPPer?
// -27
:OUTPut1:AFIXed:RANGe:LOW?
// -50
Usage:
Query only
Manual operation:
See
:OUTPut<hw>:PROTection:CLEar
Resets the protective circuit after it has been tripped.
To define the output state, use the command
.
Example:
:OUTPut1:PROT:CLE
Resets the protective circuit of the RF output.
Usage:
Event
Manual operation:
See
:OUTPut<hw>:PROTection:TRIPped?
Queries the state of the protective circuit.
Return values:
<Tripped>
1 | ON | 0 | OFF
*RST:
0
Example:
:OUTPut1:PROT:TRIP
queries the state of the protective circuit of the RF output.
Response: 0
the protective circuit has not tripped.
Usage:
Query only
Manual operation:
See
OUTPut subsystem