Interface functions
Instruments which can be controlled via GPIB-bus interface can be equipped with dif-
ferent interface functions.
lists the interface functions for the R&S
SMM100A.
Table A-2: GPIB-bus interface functions
Control character
Interface function
SH1
Handshake source function (source handshake), full capability
AH1
Handshake sink function (acceptor handshake), full capability
L4
Listener function, full capability, de-addressed by MTA.
T6
Talker function, full capability, ability to respond to serial poll, deaddressed
by MLA
SR1
Service request function (Service Request), full capability
PP1
Parallel poll function, full capability
RL1
Remote/Local switch over function, full capability
DC1
Reset function (Device Clear), full capability
DT1
Trigger function (Device Trigger), full capability