Maintenance, storage and disposal
R&S
®
SMM100A
1022
User Manual 1179.1341.02 ─ 05
16.3.3.1
Internal adjustment settings
Access:
► Select "System Config > Setup > General > Internal Adjustments".
In this dialog, you can perform internal calibration routines, and get information on
the last performed calibration.
The "Temperature Offset" indicates the deviation of the current temperature of the
instrument, compared to the temperature of the last adjustment.
The remote commands required to define these settings are described in
How to: See
Chapter 16.3.3.2, "How to use the internal adjustments"
Adjust All
Performs all available internal calibration routines of the instrument.
How to: See
Chapter 16.3.3.2, "How to use the internal adjustments"
Remote command:
Last Full Adjustment
Displays the date of the last fully performed adjustment.
Remote command:
Time
Displays the elapsed time since the last full adjustment.
If the last adjustment has been performed more than 100 days ago, the background
color of the parameter turns red.