![Rohde & Schwarz SMM100A Скачать руководство пользователя страница 875](http://html.mh-extra.com/html/rohde-and-schwarz/smm100a/smm100a_user-manual_1477939875.webp)
Remote control commands
R&S
®
SMM100A
875
User Manual 1179.1341.02 ─ 05
13.16.4.8
SOURce:BB:PROGress subsystem general commands
In the R&S
SMM100A, some calculation processes take longer time. While operating
the instrument manually, you can observe the status of an initiated process by the busy
indicator. The following commands fulfill the same task in the remote control operation.
Example: Querying the status of the Create Waveform file process
The following is an example on how to use these commands to retrieve information
about how many percent of the initiated process are completed.
[:SOURce<hw>]:BB:PROGress:MCODer?
........................................................................875
[:SOURce<hw>]:BB:PROGress:MCODer:ARBitrary:MCARrier?
..........................................875
[:SOURce<hw>]:BB:PROGress:MCODer:ARBitrary:WSEGment?
.......................................875
[:SOURce<hw>]:BB:PROGress:MCODer?
[:SOURce<hw>]:BB:PROGress:MCODer:ARBitrary:MCARrier?
[:SOURce<hw>]:BB:PROGress:MCODer:ARBitrary:WSEGment?
Queries the status of an initiated process, like for example the calculation of a signal in
accordance to a digital standard, or the calculation of a multi-carrier or multi-segment
waveform file.
Return values:
<WSegment>
integer
Indicates the task progress in percent
Range:
0 to 100
*RST:
100
Example:
See
Example "Querying the status of the Create Waveform file
Usage:
Query only
13.16.5
SOURce:CORRection subsystem
The
SOURce:CORRection
subsystem contains the commands for defining correction
values for external test assemblies.
You can acquire the correction values anytime, regardless of the modulation settings of
the generator. The correction is performed by adding the correction values to the out-
put level of the respective RF frequency.
Determine the correction values in one of the following ways:
●
Measure the RF output level at several frequency points and enter the value pairs
manually in a table
●
Connect an R&S NRP to the generator output signal and send the command
SOURce<hw>]:CORRection:CSET:DATA[:SENSor<ch>][:POWer]:SONCe
fill the table automatically.
Correction values can be stored in files with the predefined file extension
*.uco
.
SOURce subsystem