![Rohde & Schwarz SMM100A Скачать руководство пользователя страница 746](http://html.mh-extra.com/html/rohde-and-schwarz/smm100a/smm100a_user-manual_1477939746.webp)
Remote control commands
R&S
®
SMM100A
746
User Manual 1179.1341.02 ─ 05
Manual operation:
See
[:SOURce<hw>]:BBIN:MPERiod
<MPeriod>
Sets the recording duration for measuring the baseband input signal by executed
SOURce<hw>]:BBIN:ALEVel:EXECute
.
Parameters:
<MPeriod>
integer
Range:
1 to 32
*RST:
2
Default unit: s
Example:
See
Example "Applying an external digital baseband signal (Dig
Manual operation:
See
[:SOURce<hw>]:BBIN:ALEVel:EXECute
Starts measuring the input signal. The measurement estimates the crest factor, peak
and RMS level.
Example:
See
Example "Applying an external digital baseband signal (Dig
Usage:
Event
Manual operation:
See
[:SOURce<hw>]:BBIN:POWer:CFACtor
<CFactor>
Sets the crest factor of the external baseband signal.
Parameters:
<CFactor>
float
Range:
0 to 30
Increment: 0.01
*RST:
0
Default unit: dB
Example:
See
Example "Applying an external digital baseband signal (Dig
[:SOURce<hw>]:BBIN:POWer:PEAK
<Peak>
Peak level of the external baseband signal relative to full scale of 0.5 V (in terms of dB
full scale).
SOURce subsystem