Remote control commands
R&S
®
SMM100A
721
User Manual 1179.1341.02 ─ 05
Usage:
Event
Manual operation:
See
"Connect/Disconnect All Remote"
:SCONfiguration:EXTernal:PBEHaviour
<State>
If enabled, the connection to the external instruments is retained after preset (
*RST
) of
the instrument.
Parameters:
<State>
1 | ON | 0 | OFF
*RST:
n.a. (factory preset: 0)
Example:
SCONfiguration:EXTernal:IQOutput1:RCONnection:STATe?
// 1
SOURce:FREQuency:CW?
// 60000000000
SOURce1:IQ:OUTPup:ANALog:TYPE DIFFerential
SCONfiguration:EXTernal:PBEHaviour
1
*RST
SOURce:FREQuency:CW?
//
60000000000
SCONfiguration:EXTernal:IQOutput1:RCONnection:STATe?
//
1
SOURce1:IQ:OUTPup:ANALog:TYPE?
//
DIFF
SCONfiguration:EXTernal:PBEHaviour
0
*RST
SOURce:FREQuency:CW?
//
1000000000
SCONfiguration:EXTernal:IQOutput1:RCONnection:STATe?
//
0
SOURce1:IQ:OUTPut:ANALog:TYPE?
//
SING
Manual operation:
See
"Preset behavior: Keep connections to external instru-
:SCONfiguration:EXTernal:CODer<ch>:DIRection?
:SCONfiguration:EXTernal:FADer<ch>:DIRection?
:SCONfiguration:EXTernal:RF<ch>:DIRection?
:SCONfiguration:EXTernal:IQOutput<ch>:DIRection?
Queries the connector direction.
Return values:
<Direction>
NONE | IN | OUT
*RST:
IN
SCONfiguration subsystem