General instrument functions
R&S
®
SMM100A
567
User Manual 1179.1341.02 ─ 05
Sanitize
Executes the erase procedure that sanitizes the internal memory.
The sanitizing function makes sure that no user information is stored on the instrument
when it leaves the secure environment.
To apply the change: enter the security password and confirm with "Accept". Otherwise
the change has no effect.
Note:
If the instrument is subject to high security, and you have disabled volatile mode,
the internal memory holds user-data, and thus poses a security risk.
See also
Chapter 11.5, "Using the security settings"
on page 562 for more information
on the security concept.
Remote command:
:SYSTem:SECurity:SANitize[:STATe]
11.5.2.3
Manual operation security settings
Access:
► Select "System Config > Setup > Security > Security > General > Manual Opera-
tion".
The "SW Update" tab enables you to lock front panel operation and/or reading the
display.
The remote commands available to control security settings are described in:
●
Chapter 13.9, "DISPlay subsystem"
●
Chapter 13.17, "SYSTem subsystem"
Annotation Frequency
Enables the display of the currently used frequency in the status bar.
"Disabling the frequency and level indication in the status bar"
Remote command:
Annotation Amplitude
Enables the display of the currently selected level in the status bar.
Using the security settings