Getting started
R&S
®
SMM100A
27
User Manual 1179.1341.02 ─ 05
4. Tighten all screws at the rack brackets with a tightening torque of 1.2
Nm to secure
the R&S
SMM100A in the rack.
To unmount the R&S
SMM100A from a rack
1. Loosen the screws at the rack brackets.
2.
WARNING!
The R&S
SMM100A can be heavy, e.g., if fully equipped. Use a lifting
"Lifting and carrying the product"
Bring the lifting equipment to shelf height.
3. Remove the R&S
SMM100A from the rack.
4. If placing the R&S
SMM100A on a bench top again, unmount the adapter kit from
the R&S
SMM100A. Follow the instructions provided with the adapter kit.
3.1.5
Considerations for test setup
Cable selection and electromagnetic interference (EMI)
Electromagnetic interference (EMI) can affect the measurement results.
To suppress electromagnetic radiation during operation:
●
Use high-quality shielded cables, especially for the following connector types:
–
BNC
Double-shielded BNC cables.
How to:
"To connect to non-screwable connectors (BNC)"
–
SMA
Double-shielded SMA cables.
How to:
"To connect to Ref In/Ref Out (reference = 1
–
USB
Double-shielded USB cables.
How to:
Chapter 3.1.8, "Connecting USB devices"
See also
Chapter 14.6, "Measuring USB cable quality"
–
LAN
At least CAT6 STP cables.
How to:
Chapter 3.1.7, "Connecting to LAN"
●
Always terminate open cable ends.
●
Ensure that connected external devices comply with EMC regulations.
●
Use cables of the same type and equal length for connection to the I/Q and I/Q Bar
interfaces of the instrument.
●
Use the cable R&S
SMU-Z6 for connection to the Dig I/Q interfaces of the instru-
ment. The cable is available under order number 1415.0201.02.
●
Use the cable R&S
DIGIQ-HS for connection to the HS Dig I/Q interfaces of the
instrument. The cable is available under order number 3641.2948.03.
How to:
Chapter 3.1.12, "Connecting to HS Dig I/Q"