Remote control commands
R&S
®
SMM100A
782
User Manual 1179.1341.02 ─ 05
Manual operation:
See
[:SOURce<hw>]:BB:DM:PRAMp:TIME
<Time>
Sets the power ramping rise time and fall time for a burst.
Parameters:
<Time>
float
Range:
0.25 to 16
Increment: 0.01
*RST:
1
Default unit: symbol
Example:
See
Example "Enable power ramping"
Manual operation:
See
[:SOURce<hw>]:BB:DM:PRAMp:FDELay
<FDelay>
[:SOURce<hw>]:BB:DM:PRAMp:RDELay
<RDelay>
Sets the delay in the rising edge.
Parameters:
<RDelay>
float
Range:
0 to 4
Increment: 0.01
*RST:
0
Default unit: symbol
Example:
See
Example "Enable power ramping"
Manual operation:
See
[:SOURce<hw>]:BB:DM:PRAMp:ATTenuation
<Attenuation>
Sets the level attenuation for signal ranges that are flagged with level attribute
attenu-
ated
by the control signal.
Parameters:
<Attenuation>
float
Range:
0 to 50
Increment: 0.1
*RST:
15
Default unit: dB
Example:
See
Example "Enable power ramping"
Manual operation:
See
SOURce subsystem