Multi-instrument setups
R&S
®
SMM100A
475
User Manual 1179.1341.02 ─ 05
You can only activate one channel at a time. Toggle the state of the channel to test all
channels one after another.
Remote command:
[:SOURce<hw>]:BBIN:CHANnel<ch0>:BB:STATe
Aggregated Link Sample Rate
Indicates the aggregated sample rate and the maximum sample rate, where:
●
The aggregated sample rate is the sum of the sample rates of all active channels.
It cannot exceed the maximum sample rate.
●
The R&S
SMM100A supports two standard max. sample rate values.
This parameter indicates the currently selected one, where the max. sample rate is
selected depending on the capabilities of the transmitter/receiver.
For more information, see data sheet.
Remote command:
DIG IQ Auto Setting
Activates an automatic adjustment of the baseband input signal.
The R&S
SMM100A receives peak level, level and crest factor values directly from the
connected transmitter and recognizes changes automatically.
Remote command:
[:SOURce<hw>]:BBIN:DIGital:ASETting:STATe
Measurement Period
Sets the recording duration for measuring the baseband input signal by
"Auto
Level
Set".
Note:
For accurate level measurements, set the measurement period to a time value
that is long enough to capture several periods of the input signal.
Remote command:
Auto Level Set
Starts measuring the input signal for estimating the crest factor, the peak level, and the
Revel of the input signal.
See also
"Fulfilling the prerequisites for a correct signal processing of the externally
supplied digital baseband signal"
Remote command:
[:SOURce<hw>]:BBIN:ALEVel:EXECute
Digital baseband input settings