Multi-instrument setups
R&S
®
SMM100A
453
User Manual 1179.1341.02 ─ 05
Fulfilling the prerequisites for a correct signal processing of the externally sup-
plied digital baseband signal
Chapter 9.3.4.4, "Important signal parameters and interface's characteristics"
on page 451 emphasis why and which signal parameters are important for the correct
internal signal processing of externally supplied signals in the R&S
SMM100A. This
section focuses on the way these requirements are fulfilled and explains how the
instrument deals with signals deviating from the expected values:
●
Crest factor, peak level and RMS level
These values can be adjusted in one of the following ways:
–
Adjusted manually
If you have knowledge about these values, it is best to enter them directly
–
Estimated automatically by the internal measurement function
Use the function
to trigger the R&S
SMM100A to measure the
input signal, estimate the peak level and RMS level and calculate the crest fac-
tor out of them.
Tip:
For a correct signal leveling:
1) Enable an appropriate measurement time.
2) Execute one measurement for each baseband source.
3) Set the measurement period to a time value that is long enough to capture
several periods of the input signal.
–
Determined automatically
The R&S
SMM100A can receive the values from the connected transmitter.
If the external digital signal source is a second Rohde
&
Schwarz instrument, it
signals the peak level and the RMS level of the signal over the digital input
interface. To enable the R&S
SMM100A to receive these values, enable the
parameter
In the second automatic way, the R&S
SMM100A adjusts the corresponding input
fields with the measured/received values. The internal gain control mechanism
uses these received values to adjust the input signal gain to achieve an optimal
dynamic range.
●
Sample rate
External input signals with sample rates different than the system clock are resam-
pled, as illustrated on
Figure 9-3: Resampling principle of external digital baseband signals
*
= Max. sample rate depends on connected device
f
input
= Max. sample rate
●
Maximum modulation bandwidth
Control and operation of external instruments