Multi-instrument setups
R&S
®
SMM100A
466
User Manual 1179.1341.02 ─ 05
:SCONfiguration:EXTernal:CODer<ch>:REMote:INFO?
:SCONfiguration:EXTernal:IQOutput<ch>:REMote:INFO?
:SCONfiguration:EXTernal:RF<ch>:REMote:INFO?
Frontend Configuration
Requires R&S
SMM-K553.
Accesses the "RF Frontend" dialog to configure settings of the connected external
frontend.
For more information, see the document "R&S SMM-K553 Frontend Control User Man-
ual".
Initialization Sequence
Accesses the standard "File Select" dialog for loading an existing initialization file
(extension
*.iec
).
When you establish the connection to an external instrument, i.e.
SMM100A performs the following:
●
●
Sends the content of the initialization sequence file to reconfigure the external
instrument
If necessary, perform further settings and configuration of the external instruments.
Remote command:
:SCONfiguration:EXTernal:REMote:INITialization:CATalog?
:SCONfiguration:EXTernal:CODer<ch>:REMote:INITialization:FILE
:SCONfiguration:EXTernal:IQOutput<ch>:REMote:INITialization:FILE
:SCONfiguration:EXTernal:RF<ch>:REMote:INITialization:FILE
Remove Initialization Sequence
Deletes the loaded sequence.
Apply
Confirms the settings and displays the external instrument in the "External RF and I/Q"
dialog.
Remote command:
:SCONfiguration:EXTernal:CODer<ch>:REMote:ISELect
:SCONfiguration:EXTernal:IQOutput<ch>:REMote:ISELect
:SCONfiguration:EXTernal:RF<ch>:REMote:ISELect
Apply and Connect
Confirms the settings and triggers the connection establishment. In the connected
instruments, the status of this active remote connection is displayed by the "Remote"
indication.
System configuration settings