![Rohde & Schwarz NRT 1080.9506.02/.62 Скачать руководство пользователя страница 91](http://html.mh-extra.com/html/rohde-and-schwarz/nrt-1080-9506-02-62/nrt-1080-9506-02-62_operating-manual_1477860091.webp)
NRT
Structure and Syntax of the Device Messages
1080.9506.02
3.11
E-6
Boolean Parameters
Boolean parameters represent two states. The ON state (logically true) is rep-
resented by ON or a numerical value unequal to 0. The OFF state (logically
untrue) is represented by OFF or the numerical value 0. 0 or 1 is provided in a
query.
Example:
Setting command:
:CONTrol:POWer:STATe ON
Query:
:CONTrol:POWer:STATe?
Response:
1
Text
Text parameters observe the syntactic rules for key words, i.e. they can be en-
tered using a short or long form. Like any parameter, they have to be separated
from the header by a white space. In the case of a query, the short form of the
text is provided.
Example:
Setting command
:
:INPut2:PORT:POSition SOURce
Query:
:INPut2:PORT:POSition?
Response:
SOUR
Strings
Strings must always be entered within quotation marks (’ or ").
Example:
:SENSe1:FUNCtion:ON
"
POWer:FORWard:AVERage
"
or
:SENSe1:FUNCtion:ON ’POWer:FORWard:AVERage’
Block data
Block data are a transmission format which is suitable for the transmission of
large amounts of data. A command using a block data parameter has the fol-
lowing structure:
Example:
:HEADer:HEADer #45168xxxxxxxx
The ASCII character # introduces the data block. The next number indicates
how many of the following digits describe the length of the data block. In the ex-
ample the 4 following digits indicate the length to be 5168 bytes. The data bytes
follow. During the transmission of these data bytes all delimiters or other control
signs are ignored.
深圳德标仪器
135-1095-0799