IEC/IEEE Bus Interface
NRT
1080.9506.02
A.2
E-4
2. Control bus with 5 lines
IFC (Interface Clear),
active LOW resets the interfaces of the instruments connected to the default setting.
ATN (Attention),
active LOW signals the transmission of interface messages
inactive HIGH signals the transmission of device messages.
SRQ (Service Request),
active LOW enables the connected device to send a service request to the controller.
REN (Remote Enable),
active LOW permits switchover to remote control.
EOI (End or Identify),
has two functions in connection with ATN:
ATN=HIGH
active LOW marks the end of data transmission.
ATN=LOW
active LOW triggers a parallel poll.
3. Handshake bus with three lines
DAV (Data Valid),
active LOW signals a valid data byte on the data bus.
NRFD (Not Ready For Data),
active LOW signals that one of the connected devices is not ready for data transfer.
NDAC (Not Data Accepted),
active LOW signals that the instrument connected is accepting the data on the data bus.
Annex A.1.3
Interface Functions
Instruments which can be remote controlled via IEC bus can be equipped with different interface
functions. Table A-1 lists the interface functions valid for the NRT.
Table A-1
Interface functions
Control
character
Interface function
SH1
Handshake source function (source handshake)
AH1
Handshake drain function (acceptor handshake)
L4
Listener function
T6
Talker function, ability to respond to serial poll
SR1
Service request function
PP1
Parallel poll function
RL1
Remote/Local switchover function
DC1
Reset function (Device Clear)
DT1
Trigger function (Device Trigger)
深圳德标仪器
135-1095-0799