NRT
UTILity Menu
1080.9506.02
2.47
E-8
Selection
Description
Available
HIGH
If the measured value exceeds the upper scale limit, a high
signal (
>
3 V) will be output.
Always
INBND
If the measured value is within the two scale limits, a high
signal (
>
3 V) will be output.
Always
OUTBND
If the measured value is out of the two scale limits, a high
signal (
>
3 V) will be output.
Always
Preset:
HIGH
IEC/IEEE command:
:TRIGger[:TRIGger]:SOURce INTernal
:SENSE<n>:SWR:SIGNal NONe | BEEPer
:SENSe<n>:POWer:REFLection:RANGe:LIMit[:STATe] OFF
:SENSe<n>:POWer[:POWer]:RANGe:AUTO OFF
:SENSe<n>:POWer[:POWer]:RANGe:LIMit[:STATe] ON
POWER
:SENSe<n>:POWer[:POWer]:RANGe:LIMit:DETect
HIGH|INBound|OUTBound
:TRIGger[:TRIGger]:SOURce INTernal
:SENSE<n>:SWR:SIGNal NONe | BEEPer
:SENSe<n>:POWer[:POWer]:RANGe:LIMit[:STATe] OFF
:SENSe<n>:POWer:REFLection:RANGe:AUTO OFF
:SENSe<n>:POWer:REFLection:RANGe:LIMit[:STATe] ON
REFL
:SENSe<n>:POWer:REFLection:RANGe:LIMit:DETect
HIGH|INBound|OUTBound
2.5.9.3
Selection of Output Level for SWR Alarm
This menu item is used to configure the electrical output signal in the case of an SWR alarm.
UTILity
–
AUX/ IO – SWR.ALARM –
LOGIC LEVEL:
Selection
Description
Available
LOW
Outputs a LOW level in the case of an SWR alarm (< 0.8 V)
Always
HIGH
Outputs a HIGH level in the case of an SWR alarm (> 3 V).
Always
Preset:
HIGH
IEC/IEEE command:
:SENSe<n>:POWer[:POWer]:RANGe:LIMit[:STATe] OFF
:SENSe<n>:POWer:REFLection:RANGe:LIMit[:STATe] OFF
:SENSE<n>:SWR:SIGNal TTLSignal | BOTH
:SENSE<n>:SWR:SIGNal[:TTLSignal]:LEVel LOW | HIGH
深圳德标仪器
135-1095-0799
Содержание NRT 1080.9506.02/.62
Страница 2: ...深圳德标仪器 135 1095 0799 ...
Страница 4: ...深圳德标仪器 135 1095 0799 ...
Страница 6: ...深圳德标仪器 135 1095 0799 ...
Страница 16: ...深圳德标仪器 135 1095 0799 ...
Страница 18: ...深圳德标仪器 135 1095 0799 ...
Страница 178: ...深圳德标仪器 135 1095 0799 ...
Страница 190: ...深圳德标仪器 135 1095 0799 ...