SENSe Subsystem
NRT
1080.9506.02
3.28
E-6
3.5.11 SENSe
Subsystem
The SENSe subsystem provides all important commands for the configuration of the measurement (eg
test frequency, video bandwidth and reference power) and setting of the measurement function (eg
average forward power or SWR). Further command in the SENSe subsystem refer to the settings of the
bargraphs and the configuration of SWR monitoring.
The definition of the reference plane (source or load) and the assignment of the forward direction belong
to the INPut subsystem (see previous section).
All settings made in the SENSe subsystem only apply to the specified sensor with the numeric suffix
<n>:
0 :
Rear-panel connector on option NRT-B1
1 :
Front-panel connector (standard)
2, 3 :
Rear-panel connectors (option NRT-B2)
Note:
The keyword SENSe may be omitted for <n>=1 so that the following two commands are
equivalent:
:SENSe1:FUNCTion
"
POWer:FORWard:AVERage
"
FUNCTion
"
POWer:FORWard:AVERage
"
Table 3-9 SENSe subsystem
Command
Parameter
Unit
Notes
:SENSe<n>:
BANDwidth | BWIDth
:VIDeo
:FNUMber
<f>
Settings are sensor-
dependent, see table 2-1.
BURSt
:MODE
AUTO | USER
:PERiod
0 to 1 s | MINimum | MAXimum
s
:WIDTh
0 to 1 s | MINimum | MAXimum
s
DM
:STATe
ON | OFF
:STANdard
IS95 | WCDMA | DVBT | DAB
Settings are sensor-
dependent
:WCDMa
:CRATe
0...8.2 E 6 | MINimum | MAXimum |
DEFault
s
–1
Settings are sensor-
dependent
FREQuency
[:CW|:FIXED]
0 to 200 GHz | MINimum | MAXimum |
DEFault
Hz
Settings are sensor-
dependent
RRESolution
LOW | HIGH
FUNCtion
:CONCurrent
ON | OFF
:OFF
:ALL<m>
<Measurement Function>
See command description
深圳德标仪器
135-1095-0799
Содержание NRT 1080.9506.02/.62
Страница 2: ...深圳德标仪器 135 1095 0799 ...
Страница 4: ...深圳德标仪器 135 1095 0799 ...
Страница 6: ...深圳德标仪器 135 1095 0799 ...
Страница 16: ...深圳德标仪器 135 1095 0799 ...
Страница 18: ...深圳德标仪器 135 1095 0799 ...
Страница 178: ...深圳德标仪器 135 1095 0799 ...
Страница 190: ...深圳德标仪器 135 1095 0799 ...