SENSe Subsystem
NRT
1080.9506.02
3.34
E-6
:SENSe<n>:POWer:REFerence 0..100E6 W | -200..+200dBm | MINimum | MAXimum
This command enters the reference values (in W or dBm) for the relative power indication (in %,
dB, or dBm), see Section 2.3.8.1). Units: Watt or dBm.
Example:
:SENS1:POW:REF 10W
*RST value: 1W
:SENSe<n>:POWer:CCDFunction:REFerence 0 to 100E6 W | -200 to +200 dBm | MINimum |
MAXimum
This command defines the threshold for the cumulative distribution function (CCDF). The distri-
bution function states the probability (in %) of the envelope power lying above the threshold.
Units: W or dB or dBm
Example:
:SENS2:POW:CCDF:REF 10W
*RST value: sensor-dependent
:SENSe<n>:POWer[:POWer]:RANGe
This node provides all commands for setting the lefthand bargraph (power indication) and for
configuring the AUX T T L connector as an output for power monitoring.
Note:
Differing from the specifications of the SCPI Standard for RANGe commands, these
command do not set the measurement range of the NRT but only define a display or
tolerance range.
:SENSe<n>:POWer[:POWer]:RANGe:AUTO ON | OFF
This command switches the automatic adaptation of the bargraph scaling to the measured power
value on or off. If autoranging has been selected, the scale limits of the bargraphs are automati-
cally adapted to the current measured value. If autoranging is switched off, the limits and hence
the scale remain fixed.
Example:
:SENS1:POW:RANG:AUTO OFF
*RST value: ON
:SENSe<n>:POWer[:POWer]:RANGe:LIMit[:STATe] ON | OFF
This command can be used to define the AUX T T L connector as a monitoring output for the
power indication (ON state). In OFF state, the connector may be defined either as a monitoring
output for the matching or as a trigger input.
If the AUX T T L connector iss assigned more than one function, the error message –221 "Set-
tings conflict“ is generated and the first setting retained.
Example:
:TRIG:SOUR INT
:SENS1:POW:REFL:RANG:LIM OFF
:SENS1:POW:RANG:LIM ON
*RST value: OFF
:SENSe<n>:POWer[:POWer]:RANGe:LIMit:DETect INBound | OUTBound | HIGH
This command defines the conditions for a logic high level (> 2.7 V) being output at the AUX T T L
connector if this connector has been defined as a monitoring output for the power indication (eg
with the command :SENSe<n>:POWer:RANGe:LIMit ON).
OUTBound
A high level is output if the measured power is out of the rangedefined
by
the
scale limits of the lefthand bargraph.
INBound
A high level is output if the measured power is within the range specified by the
scale limits of the lefthand bargraph.
HIGH
A high level is output if the measured power exceeds the upper scale limit of the
lefthand bargraph.
Example:
:SENS1:POW:RANG:LIM:DET INB
*RST value: HIGH
深圳德标仪器
135-1095-0799
Содержание NRT 1080.9506.02/.62
Страница 2: ...深圳德标仪器 135 1095 0799 ...
Страница 4: ...深圳德标仪器 135 1095 0799 ...
Страница 6: ...深圳德标仪器 135 1095 0799 ...
Страница 16: ...深圳德标仪器 135 1095 0799 ...
Страница 18: ...深圳德标仪器 135 1095 0799 ...
Страница 178: ...深圳德标仪器 135 1095 0799 ...
Страница 190: ...深圳德标仪器 135 1095 0799 ...