NRT
CONFig Menu
1080.9506.02
2.25
E-8
a)
Entry of lower scale limit
CONFig - SCALE
-
POWER
(REFL)
-
LOWER:
Unit:
none
Preset:
0.0
IEC/IEEE command:
:SENSe<n>:POWer[:POWer]:RANGe:AUTO OFF
:SENSe<n>:POWer[:POWer]:RANGe:LOWer 1.8
In remote control mode, the bargraph for reflection indication is scaled with the keyword
:REFlection
instead of
[:POWer]
.
b)
Entry of upper scale limit:
CONFig - SCALE
-
POWER
(REFL)
-
UPPER:
Unit:
none
Preset:
1.0
IEC/IEEE command:
:SENSe<n>:POWer[:POWer]:RANGe:AUTO OFF
:SENSe<n>:POWer[:POWer]:RANGe:UPPer 2.1
For manual entry of the scale limits, the bargraph is operated in FIXED mode. The scale limits remain
fixed and are not automatically adapted. Manual return to automatic scaling is only possible via the two
scaling keys below the bargraphs (see Section 2.2.3).
The scaling of the bargraphs is dimensionless. If another unit is selected for the measured value, the
scaling is not converted.
2.3.6
RESOL (Resolution of Indication in Display)
The
CONFig - RESOL
menu provides the possibility of indicating the measurement results with low
(LOW)
or high
(HIGH)
resolution.
CONFig - RESOL:
High resolution should be selected if very small changes of the measured values are to be displayed. To
minimize flickering of the last digit, there is stronger filtering (averaging) of the measured values. The
resolution setting does not affect the output format in the remote-control mode, but it does affect the
measurement speed.
Selection
Description
Available
LOW
Indication of results with one digit less in the display.
Weaker filtering of measured values.
Always
HIGH
Indication of results with full number of digits in the display.
Stronger filtering of of measured values.
Always
Preset:
depends on the sensor used
IEC/IEEE command:
:
SENSe<n>:RRESolution LOW|HIGH
深圳德标仪器
135-1095-0799
Содержание NRT 1080.9506.02/.62
Страница 2: ...深圳德标仪器 135 1095 0799 ...
Страница 4: ...深圳德标仪器 135 1095 0799 ...
Страница 6: ...深圳德标仪器 135 1095 0799 ...
Страница 16: ...深圳德标仪器 135 1095 0799 ...
Страница 18: ...深圳德标仪器 135 1095 0799 ...
Страница 178: ...深圳德标仪器 135 1095 0799 ...
Страница 190: ...深圳德标仪器 135 1095 0799 ...