Brief Instructions
NRT
1080.9506.02
3.2
E-6
CALL IBWRT(sensor%,
"
:UNIT<n>:POW:REL:STAT ON
"
)
Activate relative units for forward
measured values
CALL IBWRT(sensor%,
"
:UNIT<n>:POW:REL PCT
"
)
Set relative units to %
CALL IBWRT(sensor%,
"
*TRG
"
)
Trigger
CALL IBRD(sensor%,measured_value)
Read measurement result
The instrument measures the average forward power of a signal and displays the deviation relative
to the reference value in units of percent.
½
To return to manual control, press the
SEL
LOCAL
key at the front panel.
3.1.2 RS-232
Interface
It is assumed that the baud rate at the NRT is set to 9600 and the handshake mode is set to XON/OFF
(factory setting, see Section 2.5.8).
½
Connect NRT and controller using the zero-modem cable (see Annex A, "Interfaces").
½
Enter the following command at the controller (under DOS) to configure the controller interface:
mode com<x>: 9600, n, 8, 1
<x>=1 or 2, according to connector used.
½
Create the following ASCII file at the controller (for <m> insert the number of the sensor port used
(0:NRT-B1, 1: front panel connector):
*RST;*CLS;*WAI
Reset instrument
:SENS<m>:FREQ DEF
Frequency response correction
:SENS<m>:POW:REF 27dBm
Set reference power to 27 dBm
:SENS<m>:FUNC
"
POW:FORW:AVER
"
Switch on average power measurement
:UNIT<m>:POW:REL:STAT ON
Activate the display of forward measured values in relative
units
:UNIT<m>:POW:REL PCT
Set relative unit to %
*TRG
Trigger
(Blank line)
½
Transfer ASCII file to unit via RS-232 interface. To this end, enter the following command at the con-
troller (under DOS):
copy <filename> com<x>:
The instrument measures the average forward power of a signal and displays the deviation relative to
the reference value in units of percent.
½
To return to manual control, press the
SEL
LOCAL
key at the front panel.
深圳德标仪器
135-1095-0799
Содержание NRT 1080.9506.02/.62
Страница 2: ...深圳德标仪器 135 1095 0799 ...
Страница 4: ...深圳德标仪器 135 1095 0799 ...
Страница 6: ...深圳德标仪器 135 1095 0799 ...
Страница 16: ...深圳德标仪器 135 1095 0799 ...
Страница 18: ...深圳德标仪器 135 1095 0799 ...
Страница 178: ...深圳德标仪器 135 1095 0799 ...
Страница 190: ...深圳德标仪器 135 1095 0799 ...