Basic Operating Functions
1080.9506.02
2.11
E-8
2.2.2
Operation via Function Keys
AVG
ENV
The
AVG / ENV
key is for switching between measurement of the average
power (
AVG)
and a configurable measurement function, the envelope pa-
rameter (
ENV).
See Section 2.3.1.
IEC/IEEE command:
Example: setting the average function:
:SENSe<n>:FUNCtion:OFF "....."
The current measurement function must be deactivated before the average
function can be switched on.
:SENSe<n>:FUNCtion[:ON]
"POWer:FORWard:AVERage"
RE F
The
∆
REF
key accepts the currently indicated measured value as reference
value. This measured value can be checked in the
CONFig
menu under the
item
SPEC
-
∆
REF
, see Section 2.3.8.1.
IEC/IEEE command:
:SENSe<n>:DATA?
<value> = response
:SENSe<n>:POWer:REFerence <value>
dBm
W
The
dBm /
∆
/ W
key
selects the unit relevant for all power indications.
The key allows selection of W, dBm or a configurable relative unit, see Sec-
tion 2.3.9.
IEC/IEEE command:
Example: selection of W:
:UNIT<n>:POWer:RELative:STATe OFF
:UNIT<n>:POWer W
SWR
RFL
The
SWR / RFL
key toggles between the possible reflection parameters.
The key allows selection of the standing wave ratio SWR or of a configur-
able display function, see Section 2.3.10.
IEC/IEEE command:
Example: selection of SWR
:SENSe<n>:FUNCtion[:ON] "POWer:S11"
:UNIT<n>:POWer:REFLection SWR
深圳德标仪器
135-1095-0799
Содержание NRT 1080.9506.02/.62
Страница 2: ...深圳德标仪器 135 1095 0799 ...
Страница 4: ...深圳德标仪器 135 1095 0799 ...
Страница 6: ...深圳德标仪器 135 1095 0799 ...
Страница 16: ...深圳德标仪器 135 1095 0799 ...
Страница 18: ...深圳德标仪器 135 1095 0799 ...
Страница 178: ...深圳德标仪器 135 1095 0799 ...
Страница 190: ...深圳德标仪器 135 1095 0799 ...