INPut Subsystem
NRT
1080.9506.02
3.26
E-6
3.5.10 INPut
Subsystem
The INPut subsystem contains all commands for defining the reference plane and the forward direction
for directional power sensors. All other settings referring to the measurement task are made in the
SENSe subsystem.
The INPUT subsystem is only available for direction sensors with directional coupler. The sensor is
selected by defining the numeric suffix <n>.
Table 3-8
INPut subsystem
Command
Parameter
Unit
Notes
:INPut<n>
:PORT
:POSition
SOURce | LOAD
:OFFset
0 dB...100 dB | MINimum | MAXimum | DEFault
dB
:SOURce
1 | 2 | DEFault
:AUTO
ON | OFF
<n>
³
0..3 (sensor port)
:INPut<n>:PORT:POSition SOURce | LOAD
This command selects the reference plane. It defines the sensor port to which the measurement
results are to be referred to.
SOURce
referred to source.
LOAD
referred to load.
Example:
:INP2:PORT:POS SOUR
*RST value: sensor-dependent
:INPut<n>:PORT:OFFSet 0...100 dB | MINimum | MAXimum | DEFault
This command allows the transmission loss in a cable connecting the desired measurement point
and the sensor to be taken into account. If the measurement position is set to LOAD
(:INPut<n>:PORT:POSition LOAD) the cable is assumed to be connected between the sensor
and the load, otherwise between the sensor and the source.
Example:
:
INP1:PORT:OFFS 1.25
*RST value: sensor-dependent
:INPut<n>:PORT:SOURce 1 | 2 | DEFault
This command has to be used for defining the forward direction if the automatic assignment has
been switched off with the command INPut:PORT:SOUR:AUTO OFF (see below).
The forward direction is defined by stating the source RF connector (1, 2 or DEFault). For sen-
sors with hardware-defined privileged direction, the default setting causes selection of exactly this
privileged direction as forward direction (eg NRT-Z44). Otherwise the DEFault setting is identical
with the selection of port 1 as the source RF connector. The numbers 1 and 2 are clearly shown
on the case or type label of each directional sensor.
Example:
:INP2:PORT:SOUR 1
*RST value: sensor-dependent
深圳德标仪器
135-1095-0799
Содержание NRT 1080.9506.02/.62
Страница 2: ...深圳德标仪器 135 1095 0799 ...
Страница 4: ...深圳德标仪器 135 1095 0799 ...
Страница 6: ...深圳德标仪器 135 1095 0799 ...
Страница 16: ...深圳德标仪器 135 1095 0799 ...
Страница 18: ...深圳德标仪器 135 1095 0799 ...
Страница 178: ...深圳德标仪器 135 1095 0799 ...
Страница 190: ...深圳德标仪器 135 1095 0799 ...