NRT
IEC/IEEE Bus Interface
1080.9506.02
A.3
E-4
Annex A.1.4
Interface Messages
Interface messages are transmitted to the instrument on the data lines, with the attention line being
active (LOW). They serve to communicate between controller and instrument.
Universal Commands
Universal commands are encoded in the range 10 through 1F hex. They are effective for all instruments
connected to the bus without pervious addressing.
Table A-2
Universal Commands
Command
QuickBASIC command
Effect on the instrument
DCL
(Device Clear)
IBCMD (controller%, CHR$(20))
Aborts processing of the commands just received
and sets the command processing software to a
defined initial state. Does not change the instrument
setting.
IFC
(Interface Clear)
IBSIC (controller%)
Resets the interfaces to the default setting.
LLO
(Local Lockout)
IBCMD (controller%, CHR$(17))
The LOC/IEC ADDR key is disabled.
SPE
(Serial Poll Enable)
IBCMD (controller%, CHR$(24))
Ready for serial poll.
SPD
(Serial Poll Disable)
IBCMD (controller%, CHR$(25))
End of serial poll.
PPU
(Parallel Poll
Unconfigure)
IBCMD (controller%, CHR$(21))
End of the parallel-poll state.
Addressed Commands
Addressed commands are encoded in the range 00 through 0F hex. They are only effective for
instruments addressed as listeners.
Table A-3
Addressed Commands
Command
QuickBASIC command
Effect on the instrument
SDC
(Selected Device
Clear)
IBCLR (device%)
Aborts processing of the commands just received
and sets the command processing software to a
defined initial state. Does not change the instrument
setting.
GET
(Group Execute
Trigger)
IBTRG (device%)
Triggers a previously active device function. The
effect of the command is the same as with that of a
pulse at the external trigger signal input.
GTL
(Go to Local)
IBLOC (device%)
Transition to the "Local" state (manual control).
PPC
(Parallel Poll
Configure)
IBPPC (device%, data%)
Configure instrument for parallel poll. The
QuickBASIC command additionally executes PPE /
PPD.
深圳德标仪器
135-1095-0799
Содержание NRT 1080.9506.02/.62
Страница 2: ...深圳德标仪器 135 1095 0799 ...
Страница 4: ...深圳德标仪器 135 1095 0799 ...
Страница 6: ...深圳德标仪器 135 1095 0799 ...
Страница 16: ...深圳德标仪器 135 1095 0799 ...
Страница 18: ...深圳德标仪器 135 1095 0799 ...
Страница 178: ...深圳德标仪器 135 1095 0799 ...
Страница 190: ...深圳德标仪器 135 1095 0799 ...