SENSe Subsystem
NRT
1080.9506.02
3.32
E-6
:SENSe<n>:FUNCtion:OFF <Measurement Function>
This command switches off a specified measurement function. If the command is in the form of a
query, the headers of all switched-off measurement functions will be returned in the sequence
defined under [SENS<n>:]FUNC:ON.
The command is a function call or query and therefore has no *RST value.
Example:
:SENS2:FUNC:OFF
"
POW:REV
"
:SENS2: FUNC:OFF?
The following response would be possible:
"
POW:CFAC
"
,
"
POW:FORW:AVER:BURS
"
,
"
POW:FORW:PEP
"
,
"
POW:FORW:CCDF
"
,
"
POW:ABS:AVER
"
,
"
POW:ABS:AVER:BURS
"
,
"
POW:ABS:PEP
"
,
"
POW:REV
"
:SENSe<n>:FUNCtion:OFF:ALL<m>
This command switches off all measurement functions referring to a specific channel no. <m>, i. e.,
...ALL1 switches off all forward measurement functions, ...ALL2 switches off all measurement func-
tions in the reverse channel.
The command is a function call and therefore has no *RST-value.
Example:
:SENS1:FUNC:OFF:ALL1
:SENSe<n>:FUNCtion[:ON] <Measurement Function>
This command switches on a specified measurement function. If this command is in the form of a
query, the headers of all switched-on measurement functions (maximum of two functions) are
returned.
The command is a function call or query and therefore has no *RST value.
The following measurement functions are available:
"
POWer:CFACtor
"
Crest Factor
Sensor
"
POWer:FORWard:AVERage
"
Average power (forward)
Always
"
POWer:FORWard:AVERage:BURSt
"
Average burst power (forward)
Always
"
POWer:FORWard:PEP
"
Peak envelope power (forward)
Sensor
"
POWer:FORWard:CCDFunction
"
Cumulative distribution function of PEP
Sensor
"
POWer:ABSorbtion:AVERage
"
Average absorbed power (F-R)
Sensor
"
POWer:ABSorbtion:AVERage:BURSt
"
Average absorbed burst power (F-R)
Sensor
"
POWer:ABSorbtion:PEP
"
Absorbed peak envelope power (F-R)
Sensor
"
POWer:REVerse
"
Reverse power
Sensor
"
POWer:S11
"
|
"
POWer:REFLection
"
Matching of load. The UNIT system
defines whether the output is in form of
SWR, return loss, reflection coefficient,
or power ratio R/F.
Sensor
A query returns the responses in the above sequence.
Example:
:SENS2:FUNC
"
POW:REV
"
:SENS2:FUNC?
A possible response is:
"
POW:FORW:AVER
"
,
"
POW:REV
"
深圳德标仪器
135-1095-0799
Содержание NRT 1080.9506.02/.62
Страница 2: ...深圳德标仪器 135 1095 0799 ...
Страница 4: ...深圳德标仪器 135 1095 0799 ...
Страница 6: ...深圳德标仪器 135 1095 0799 ...
Страница 16: ...深圳德标仪器 135 1095 0799 ...
Страница 18: ...深圳德标仪器 135 1095 0799 ...
Страница 178: ...深圳德标仪器 135 1095 0799 ...
Страница 190: ...深圳德标仪器 135 1095 0799 ...