NRT
RS-232-C Interface
1080.9506.02
A.5
E-4
Annex A.2.3
Transmission parameters
In order to ensure error-free and correct data transmission, the parameters of the instrument and the
controller must be set identically. The parameters are defined in the UTILITIES-SYSTEM-RS232 menu.
Transmission rate
Four different baud rates can be set in the instrument:
(baud rate)
1200, 2400, 4800, 9600.
Data bits
Data transmission is in 8-bit ASCII code. The first bit transmitted
is the LSB (Least Significant Bit).
Start bit
Each data byte begins with a start bit. The falling edge of the start
bit indicates the beginning of the data byte.
Parity bit
No parity bit is used.
Stop bit
The transmission of a data byte is terminated by a stop bit.
Example:
Transmission of character A (41 hex) in the 8-bit ASCII code.
01
02
03
04
05
06
07
08
09
10
Bit 01 = Start bit
Bit 02...09 = Data bits
Bit 10 = Stop bit
Bitduration= 1/baud rate
Annex A.2.4
Interface functions
For interface control, some control characters from the ASCII code range of 0 to 20 hex are predefined
and can be transmitted via the interface (see Table A-4).
Table A-4
Control strings or control characters of the RS-232-C interface
Control Character
Function
Break (at least 1 character only log 0)
Reset instrument
0Dhex, 0Ahex
Terminator <CR>, <LF>
Switchover between local / remote
深圳德标仪器
135-1095-0799
Содержание NRT 1080.9506.02/.62
Страница 2: ...深圳德标仪器 135 1095 0799 ...
Страница 4: ...深圳德标仪器 135 1095 0799 ...
Страница 6: ...深圳德标仪器 135 1095 0799 ...
Страница 16: ...深圳德标仪器 135 1095 0799 ...
Страница 18: ...深圳德标仪器 135 1095 0799 ...
Страница 178: ...深圳德标仪器 135 1095 0799 ...
Страница 190: ...深圳德标仪器 135 1095 0799 ...