NRT
CALCulate Subsystem
1080.9506.02
3.21
E-6
3.5.6 CALCulate
Subsystem
The CALCulate subsystem provides the commands for determining the min/max values. All settings
refer to the sensor addressed (suffix <n>, see Section 3.5.1).
Table 3-4
CALCulate subsystem
Command
Parameter
Unit
Notes
:CALCulate<n>
:LIMit
[:STATe]
:TYPE
ON | OFF
MINimum | MAXimum | DIFFerence
<n> = 0 to 3
:CALCulate<n>:LIMit
This node controls the determination of min/max limit values of the NRT.
:CALCulate<n>:LIMit[:STATe] ON | OFF
These commands are used to start (ON) or stop (OFF) the search for the min/max values, which
covers the measurement functions set under subsystem :SENSE<n>: FUNCtion [:ON] <measure-
ment function>. The min/max memories are automatically reset upon start. During a search for the
max/min values, the value set under subsystem :CALCulate<n>:LIMit:TYPE (maximum, minimum or
the difference thereof) is returned to a request for data, not however the currently measured value.
The request for data is in the form of the query :SENSE<n>: DATA?.
Example:
:CALC1:LIM ON
*RST value: OFF
:CALCulate<n>:LIMit:TYPE MINimum | MAXimum | DIFFerence
These commands allow selection of the value to be output:
MINimum
Minimum value
MAXimum
Maximum value
DIFFerence
Difference between maximum and minimum value
Example:
:CALC1:LIM:TYPE MAX
*RST value:MAX
深圳德标仪器
135-1095-0799
Содержание NRT 1080.9506.02/.62
Страница 2: ...深圳德标仪器 135 1095 0799 ...
Страница 4: ...深圳德标仪器 135 1095 0799 ...
Страница 6: ...深圳德标仪器 135 1095 0799 ...
Страница 16: ...深圳德标仪器 135 1095 0799 ...
Страница 18: ...深圳德标仪器 135 1095 0799 ...
Страница 178: ...深圳德标仪器 135 1095 0799 ...
Страница 190: ...深圳德标仪器 135 1095 0799 ...