NRT
CONFig Menu
1080.9506.02
2.23
E-8
2.3.2
POWER (Configuration of Power Display)
With the
CONFig - POWER
menu either the forward power
(FWD)
or the net transmitted power (forward
power - reverse power,
F-R
) can be displayed. In case of a load with good matching, this makes little
difference.
CONFig - POWER:
Selection
Description
Available
FWD
Indication of forward power in the left half of the display. The
symbol
FWD
appears in the lower part of the display.
Always
F–R
Indication of the difference between forward and reverse
power in the left half of the display. This display mode is
indicated by the symbol
F–R
.
Sensor
Preset:
FWD
IEC/IEEE command:
:SENSe<n>:FUNCtion[:ON] "POWer:FORWard:..."
(FWD)
:SENSe<n>:FUNCtion[:ON] "POWer:ABSorption:..."
(F-R)
2.3.3
MAXHOLD (min/max display on/off)
The NRT is able to store the maximum and minimum values as well as the differences thereof for all
measurement functions. Two menus are available for this purpose. Via the
CONFig - MAXDISP
menu
(see following section) display of the maximum, minimum or difference value can be selected. The
menu
CONFig-MAXHOLD:
allows start and stop of the max/min determination for both the power and reflection indication.
Determination of the max/min value is started by selecting the menu item
RUN.
As soon as a value is
measured that is greater than the stored maximum value or smaller than the stored minimum value, the
memory is updated and the displayed value corrected accordingly. After the start, the max/min memory
is initialized by the first measured values.
The NRT has two sets of max/min memories, one for the power indication and one for the reflection
indication. If during a max/min determination the measurement function is changed (eg from AVG to
PEP), the max/min memories concerned (here: for power indication) are reinitialized and the max/min
determination is started for the new measurement function.
Selection
Description
Available
RUN
Start of max/min display
Always
OFF
Stop of max/min display and return to normal measure-
ment mode
Always
Preset:
OFF
IEC/IEEE command:
:CALCulate<n>:LIMit[:STATe] ON|OFF
Start/stop min/max
determination
深圳德标仪器
135-1095-0799
Содержание NRT 1080.9506.02/.62
Страница 2: ...深圳德标仪器 135 1095 0799 ...
Страница 4: ...深圳德标仪器 135 1095 0799 ...
Страница 6: ...深圳德标仪器 135 1095 0799 ...
Страница 16: ...深圳德标仪器 135 1095 0799 ...
Страница 18: ...深圳德标仪器 135 1095 0799 ...
Страница 178: ...深圳德标仪器 135 1095 0799 ...
Страница 190: ...深圳德标仪器 135 1095 0799 ...