NRT
CALibration Subsystem
1080.9506.02
3.23
E-6
:CALibration0:STATe<m> ON | OFF
This command activates the relevant calibration data set (ON) for the frequency response correction.
The command must be sent to the NRT prior to the entry of the test frequency (see Section 2.4.1.2,
Entry of Carrier Frequency for NAP Sensors). Only one calibration data set shall be active, all others
must be switched off (OFF). Entry of the calibration data is described further below.
Example:
:CAL0:STAT1 OFF
*RST value: OFF
:CAL0:STAT3 OFF
:CAL0:STAT2 ON
:SENS0:FREQ 1.2GHz
:CALibration0:FREQuency<m>:DATA <Frequency Array>
This command allows the frequencies for the individual calibration points to be entered. The fre-
quencies shall be entered as a parameter list in the sequence specified by the manufacturer, ie
starting with the lowest value. The individual frequencies are to be separated by a comma.
Example:
:CAL0:FREQ2:DATA 0.9E9,1.0E9,...,2.0E9
No *RST value
:CALibration0:LOAD<m>:DATA <Calibration Factor Array CF12>
This command is used to transfer the calibration factors for measurement direction 1
→
2 (CF12) into
the selected calibration data set. The sequence should be in accordance with the calibration fre-
quencies. In the following example the calibration factor of 96.5 % would refer to the frequency of 1
GHz from the previous example:
Example:
:CAL0:LOAD2:DATA 95.2,96.5,...,104.2
No *RST value
:CALibration0:SOURce<m>:DATA <Calibration Factor Array CF21>
This command is used to transfer the calibration factors for measurement direction 2
→
1 (CF21) into
the selected calibration data set. The sequence should be in accordance with the calibration fre-
quencies.
Example:
:CAL0:SOUR2:DATA 98.4,99.8,...,98.1
No *RST value
深圳德标仪器
135-1095-0799
Содержание NRT 1080.9506.02/.62
Страница 2: ...深圳德标仪器 135 1095 0799 ...
Страница 4: ...深圳德标仪器 135 1095 0799 ...
Страница 6: ...深圳德标仪器 135 1095 0799 ...
Страница 16: ...深圳德标仪器 135 1095 0799 ...
Страница 18: ...深圳德标仪器 135 1095 0799 ...
Страница 178: ...深圳德标仪器 135 1095 0799 ...
Страница 190: ...深圳德标仪器 135 1095 0799 ...