CONFig Menu
NRT
1080.9506.02
2.18
E-8
AV.BRST
Average Burst
... designates the average power
within
a burst. It is de-
rived by the NRT from the average power and the duty
cycle. In a submenu it must be defined whether the duty
cycle is to be determined automatically or be derived from
the values entered for burst duration and period.
Sensor
CCDF
Complementary Cumulative Distribution Function
... provides information about the probability of the meas-
ured values to lie above a specified threshold.
Sensor
Preset:
AV.BRST
IEC/IEEE command:
:SENSe<n>:FUNCtion[:ON] <Measurement Function>
2.3.1.1
Measurement of Peak Envelope Power
For measuring the peak envelope power, the video bandwidth used for measuring the detected RF sig-
nal has to be specified. The set bandwidths are sensor-dependent and indicated in the menu in plain
text as soon as the setting
PEP
is confirmed in the calling menu. The bandwidths shown in the figure
below – 4 kHz, 200 kHz, and
"FULL"
– can be set with sensors NRT-Z43/44.
"FULL"
means that the
maximum bandwidth of the sensor, i.e. approx. 4 MHz, is used.
CONFig - ENV KEY -
PEP -
VID.BW:
The video bandwidth set in this menu overwrites the settings made in the menus
CF, AV.BRST - AUTO
and
CCDF
.
IEC/IEEE command:
:SENSe<n>:FUNCtion[:ON] "POWer:FORWard:PEP"
:SENSe<n>:BANDwidth:VIDeo:FNUMber <f>
<f>
: see Table 2-1, the preset value depends on the sensor used.
PEP
Time
0
PEP
PEP
a) "FULL"
b) 200 kHz
c) 4 kHz
Po
we
r
Fig. 2-4
Effect of video bandwidth on indicated peak envelope power
The video bandwidth should be selected as narrow as possible for the measurement task at hand, since
in this way the noise superimposed on the rectified signal can be kept low. Low noise allows the meas-
urement of smaller peak envelope powers and causes less variations in indication. Reducing the band-
width too much, however, may lead to an unwanted distortion of the envelope, see Fig. 2-4.
深圳德标仪器
135-1095-0799
Содержание NRT 1080.9506.02/.62
Страница 2: ...深圳德标仪器 135 1095 0799 ...
Страница 4: ...深圳德标仪器 135 1095 0799 ...
Страница 6: ...深圳德标仪器 135 1095 0799 ...
Страница 16: ...深圳德标仪器 135 1095 0799 ...
Страница 18: ...深圳德标仪器 135 1095 0799 ...
Страница 178: ...深圳德标仪器 135 1095 0799 ...
Страница 190: ...深圳德标仪器 135 1095 0799 ...