UTILity Menu
NRT
1080.9506.02
2.46
E-8
OUT
Control connector: A logic low/high signal is output at this
connector if the measured power or SWR values are within
or out of the range limits defined for the bargraphs. In a
submenu either the lefthand (power) or the righthand bar-
graph (SWR) can be selected and the response exactly
defined.
Always
SWR.ALARM
Outputs a TTL signal with selectable level (see description
of submenu) if the conditions for a SWR warning are ful-
filled (see section 2.3.8.2).
The input/output signals at the AUX connector are TTL-compatible.
Preset:
OFF
IEC-Befehl:
:SENSe<n>:POWer[:POWer]:RANGe:LIMit[:STATe] OFF
OFF
:SENSe<n>:POWer:REFLection:RANGe:LIMit[:STATe] OFF
:SENSE<n>:SWR:SIGNal NONe | BEEPer
:TRIGger:SOURce INTernal
:SENSe<n>:POWer[:POWer]:RANGe:LIMit[:STATe] OFF
EXT.TRIG
:SENSe<n>:POWer:REFLection:RANGe:LIMit[:STATe] OFF
:SENSE<n>:SWR:SIGNal NONe | BEEPer
:TRIGger:SOURce EXTernal
The remote-control commands for the OUT and SWR ALARM modes are
described in the following sections.
2.5.9.1
Configuration as a Monitoring Output: Selection of Bargraph
In the
OUT
menu the bargraph to which the output signal is to be referred is selected.
UTILity
–
AUX/ IO – OUT:
Selection
Description
Available
POWER
Monitoring is selected for the lefthand bargraph
Always
REFL
Monitoring is selected for the righthand bargraph
Sensor
Preset:
none (last setting)
2.5.9.2
Entering the Monitoring Criteria
After selecting the bargraph the NRT requests the entry of monitoring criteria. The selection is equal for
both bargraphs.
With the parameters
HIGH, INBND
and
OUTBND
the conditions are defined under which the test signal
generates a high level (> 3 V) at the AUX connector. If the conditions are not fulfilled, low level (< 0.8 V)
will be output.
UTILity
–
AUX/ IO – OUT –
POWER (REFL):
深圳德标仪器
135-1095-0799
Содержание NRT 1080.9506.02/.62
Страница 2: ...深圳德标仪器 135 1095 0799 ...
Страница 4: ...深圳德标仪器 135 1095 0799 ...
Страница 6: ...深圳德标仪器 135 1095 0799 ...
Страница 16: ...深圳德标仪器 135 1095 0799 ...
Страница 18: ...深圳德标仪器 135 1095 0799 ...
Страница 178: ...深圳德标仪器 135 1095 0799 ...
Страница 190: ...深圳德标仪器 135 1095 0799 ...