NRT
SENSe Subsystem
1080.9506.02
3.31
E-6
:SENSe<n>:DM
This node provides the commands for correction of the measurement value for modulated signals
(see Section 2.4.3).
:SENSe<n>:DM:STATe ON | OFF
This command switches the correction of the measurement value for modulated signals on or off.
Selection of a communication standard (see below) is enabled only if the correction is switched
on.
Example:
:
SENS1:DM:STAT ON
*RST value: OFF
:SENSe<n>:DM:STANdard IS95 | WCDMa | DVBT | DAB
This command selects the communication standard. Which settings are available depends on the
sensor used. The settings are valid only if the modulation correction is switched on via the
:INPut<n>:DM:STATe ON command.
Example:
:SENS1:DM:STAN IS95
*RST value: none (last setting)
:SENSe<n>:DM:WCDMa:CRATe 0..8.2E6 | MINimum | MAXimum | DEFault
This command defines the chip rate for the communication standard WCDMA. The allowed range
is sensor-dependent; the value is always entered without the unit (s
-1
).
Example:
:SENS1:DM:WCDM:CRAT 4.096E6
*RST value: sensor-dependent
:SENSe<n>:FREQuency[:CW|:FIXED] 0 to 200 GHz | MINimum | MAXimum | DEFault
This command informs the NRT of the carrier frequency of the specified sensor.
The entered frequency value is required to correct the frequency-dependent response of the sen-
sor. For this purpose NRT sensors contain a ROM with all sensor-specific data. For NAP sensors
frequency-dependent calibration factors can be entered via the CALibration system (see Section
3.5.7, CALibration Subsystem).
Example:
:SENS1:FREQ 10 MHz
*RST value: DEF (sensor-dependent)
:SENSe<n>:RRESolution LOW | HIGH
This command changes the measurement resolution. This has an effect on the accuracy of the
measurement, its duration, and on the number of digits indicated in the display.
Example:
:SENS1:RRES LOW
*RST value: sensor-dependent
:SENSe<n>:FUNCtion
This node provides all measurement functions.
:SENSe<n>:FUNCtion:CONCurrent ON | OFF
This command defines whether several measurement functions may be simultaneously activated.
OFF
Only a single function can be active. If a new measurement function is switched on,
the previously active function is switched off automatically.
ON
Two measurement functions may be simultaneously active. If two measurement
functions are mutually exclusive, the error message. –221 "Settings conflict“ is gen-
erated and the first setting retained. All POW:FORW and POW:ABS functions are
mutually exclusive. This applies also to the POW:REV and POW:S11 functions.
Example:
:SENS2:FUNC:CONC ON
*RST value: ON
深圳德标仪器
135-1095-0799
Содержание NRT 1080.9506.02/.62
Страница 2: ...深圳德标仪器 135 1095 0799 ...
Страница 4: ...深圳德标仪器 135 1095 0799 ...
Страница 6: ...深圳德标仪器 135 1095 0799 ...
Страница 16: ...深圳德标仪器 135 1095 0799 ...
Страница 18: ...深圳德标仪器 135 1095 0799 ...
Страница 178: ...深圳德标仪器 135 1095 0799 ...
Страница 190: ...深圳德标仪器 135 1095 0799 ...