CMU-K20/K21/K22/K23/K24
Non Signalling: Power Measurements
1115.6088.12 4.19
E-11
Info
Box
The hotkey
Info Box
switches the info boxes for all displayed timeslots on or off.
For a description of the info boxes see section
P/t Multislot
on p.
4.23
ff.
Remote control
No command, display configuration only
Timing
Offset
The
Timing Offset
hotkey shifts the burst by the entered number of symbol periods,
e.g. to compensate for a known, constant timing error of the MS. See section
Measurement Control (Power Configuration – Control)
on p.
4.28
ff.
Remote control
CONFigure:POWer:MSLot:TOFFset
Modulation
View
The hotkey
Modulation View
defines the expected modulation scheme in all four
timeslots that can be measured and adjusts the power/time template. To obtain a
valid measurement result, the actual modulation in all measured slots (see
Fig. 4-3
on p.
4.18
) must be compatible with the
Modulation View
settings. See section
Measurement Control (Power Configuration – Control)
on p.
4.28
ff.
Remote control
CONFigure:POWer:MSLot:MVIew
Level
Scale
The
Level Scale
hotkey defines a maximum RF level in dBm
(Max)
and a level
range in dB
(Span)
that will be displayed in the graphical diagram and thus
determines the scale of the y axis.
Remote control
No command, display configuration only
Time
Scale
The
Time Scale
hotkey defines the start time
(Start)
and the total time interval
(Span)
that will be displayed in the graphical diagram and thus determines the
scale of the x axis.
Both values are expressed in symbol periods. 1 symbol corresponds to approx.
3.69
µ
s so that 1 timeslot comprises 156 ¼ symbols.
Start
is expressed relative to
symbol 0 of the measured timeslot (see
Fig. 4-3
on p.
4.18
).
Time Scale
only
configures the diagram; it does not affect the number of timeslots actually
measured but is modified as this number is changed (see
Slot Count
hotkey on
p.
4.57
).
Remote control
No command, display configuration only
Default
Scale
The
Default Scale
hotkey sets a default
Level Scale
and a default
Time Scale
, the
latter corresponding to the number of timeslots measured (see
Slot Count
hotkey
on p.
4.57
) plus an appropriate display margin.
Remote control
No command, display configuration only
深圳德标仪器
135-1095-0799
Содержание CMU-K20
Страница 4: ...深圳德标仪器 135 1095 0799 ...
Страница 8: ...深圳德标仪器 135 1095 0799 ...
Страница 10: ...深圳德标仪器 135 1095 0799 ...
Страница 12: ...深圳德标仪器 135 1095 0799 ...
Страница 18: ...深圳德标仪器 135 1095 0799 ...
Страница 24: ...深圳德标仪器 135 1095 0799 ...
Страница 26: ...深圳德标仪器 135 1095 0799 ...
Страница 34: ...深圳德标仪器 135 1095 0799 ...
Страница 58: ...深圳德标仪器 135 1095 0799 ...
Страница 60: ...深圳德标仪器 135 1095 0799 ...
Страница 78: ...深圳德标仪器 135 1095 0799 ...
Страница 286: ...深圳德标仪器 135 1095 0799 ...
Страница 288: ...深圳德标仪器 135 1095 0799 ...
Страница 304: ...深圳德标仪器 135 1095 0799 ...
Страница 632: ...深圳德标仪器 135 1095 0799 ...