CMU-K20/K21/K22/K23/K24
Measurement Groups (Non Signalling only)
1115.6088.12 6.29
E-12
Measured Values – Subsystem POWer:MPR
The subsystem
POWer:MPR
determines and returns the results of the combined power and modulation
measurement.
READ[:SCALar]:POWer[:NORMal][:GMSK]:MPR?
Scalar Results:
Start single shot measurement and return results
FETCh[:SCALar]:POWer[:NORMal][:GMSK]:MPR?
Read out measurement results (unsynchronized)
SAMPle[:SCALar]:POWer[:NORMal][:GMSK]:MPR?
Read out measurement results (synchronized)
Returned values
Value range
Def. value
Def. unit
FW vers.
AvgBurstPwCurr,
PeakBurstPwCurr,
BurstsOutOfTol,
BurstMatching,
–137 dBm to +53 dBm
–137 dBm to +53 dBm
0.0 % to 100.0 %
INV | MATC | NMAT | OUT | NTR | NRAM |
OFLW | UFLW | NTSC | OFF
NAN
NAN
NAN
INV
dBm
dBm
%
–
V2.00
PhErrPeakCurrent,
PhErrPeakAverage,
PhErrPeakMaxMin,
PhErrRMSCurrent,
PhErrRMSAverage,
PhErrRMSMaxMin,
FreqErrCurrent,
FreqErrAverage,
FreqErrMaxMin,
AvgBurstPwAvg
–100.0 ° to +100.0 °
–100.0 ° to +100.0 °
–100.0 ° to +100.0 °
–100.0 ° to +100.0 °
–100.0 ° to +100.0 °
–100.0 ° to +100.0 °
–1000.0 Hz to + 1000.0 Hz
–1000.0 Hz to + 1000.0 Hz
–1000.0 Hz to + 1000.0 Hz
–137 dBm to +53 dBm
NAN
NAN
NAN
NAN
NAN
NAN
NAN
NAN
NAN
NAN
deg
deg
deg
deg
deg
deg
Hz
Hz
Hz
dBm
Description of command
These commands are always queries. They start a combined power vs. time and modulation measurement and
return all scalar measurement results. For detailed information refer to the description of the analogous com-
mands in the POWer and MODulation systems.
READ:ARRay:POWer[:NORMal][:GMSK]:MPR[:CURRent]?
Traces
READ:ARRay:POWer[:NORMal][:GMSK]:MPR:AVERage?
READ:ARRay:POWer[:NORMal][:GMSK]:MPR:MAXimum?
READ:ARRay:POWer[:NORMal][:GMSK]:MPR:MINimum?
Start single shot
measurement and return results
⇒
RUN
FETCh:ARRay:POWer[:NORMal][:GMSK]:MPR[:CURRent]?
FETCh:ARRay:POWer[:NORMal][:GMSK]:MPR:AVERage?
FETCh:ARRay:POWer[:NORMal][:GMSK]:MPR:MAXimum?
FETCh:ARRay:POWer[:NORMal][:GMSK]:MPR:MINimum?
Read meas. results (unsynchronized)
⇒
RUN
SAMPle:ARRay:POWer[:NORMal][:GMSK]:MPR[:CURRent]? SAM-
Ple:ARRay:POWer[:NORMal][:GMSK]:MPR:AVERage?
SAMPle:ARRay:POWer[:NORMal][:GMSK]:MPR:MAXimum?
SAMPle:ARRay:POWer[:NORMal][:GMSK]:MPR:MINimum?
Read results (synchronized)
⇒
RUN
深圳德标仪器
135-1095-0799
Содержание CMU-K20
Страница 4: ...深圳德标仪器 135 1095 0799 ...
Страница 8: ...深圳德标仪器 135 1095 0799 ...
Страница 10: ...深圳德标仪器 135 1095 0799 ...
Страница 12: ...深圳德标仪器 135 1095 0799 ...
Страница 18: ...深圳德标仪器 135 1095 0799 ...
Страница 24: ...深圳德标仪器 135 1095 0799 ...
Страница 26: ...深圳德标仪器 135 1095 0799 ...
Страница 34: ...深圳德标仪器 135 1095 0799 ...
Страница 58: ...深圳德标仪器 135 1095 0799 ...
Страница 60: ...深圳德标仪器 135 1095 0799 ...
Страница 78: ...深圳德标仪器 135 1095 0799 ...
Страница 286: ...深圳德标仪器 135 1095 0799 ...
Страница 288: ...深圳德标仪器 135 1095 0799 ...
Страница 304: ...深圳德标仪器 135 1095 0799 ...
Страница 632: ...深圳德标仪器 135 1095 0799 ...