Non Signalling: Spectrum Measurements
CMU-K20/K21/K22/K23/K24
1115.6088.12 4.66
E-11
Fig. 4-30 Spectrum Configuration – Limits
Default Settings
The
Default All Settings
switch assigns default values to all settings in the
Limits
tab
(the default values are quoted in the command description in chapter 6 of this
manual). In addition, default switches for the individual spectrum types are
provided.
Remote control
CONFigure:SPECtrum:SWITching[:GMSK]:LIMit:LINE:DEFault ON|OFF
Limit Values
The
Limit Values
table section defines upper limits for the absolute output power of
the mobile station, measured at zero frequency span and with a filter bandwidth of
30 kHz.
Enable
Switches the limit check at the frequency or power level on and off
Power Lvl.
User-defined MS output power level (not necessarily identical with the
GSM power control levels). The CMU offers considerable flexibility with
regard to the limit line definition: They are specified at four fixed, GSM-
specific frequency offsets and up to 10 arbitrary MS power levels (see
Table 4-3
). For measured MS powers between the power levels, the
limits are determined by linear interpolation.
Remote control
CONFigure:SPECtrum:SWITching[:GMSK]:LIMit:LINE
[:ASYMmetrical]:UPPer<nr>
<PowLvl>,<Value1>,<Value2>,<Value3>,<Value4>,<Enable>
CONFigure:SPECtrum:SWITching[:GMSK]:LIMit:LINE:MODE[:UPPer]
ON|OFF
深圳德标仪器
135-1095-0799
Содержание CMU-K20
Страница 4: ...深圳德标仪器 135 1095 0799 ...
Страница 8: ...深圳德标仪器 135 1095 0799 ...
Страница 10: ...深圳德标仪器 135 1095 0799 ...
Страница 12: ...深圳德标仪器 135 1095 0799 ...
Страница 18: ...深圳德标仪器 135 1095 0799 ...
Страница 24: ...深圳德标仪器 135 1095 0799 ...
Страница 26: ...深圳德标仪器 135 1095 0799 ...
Страница 34: ...深圳德标仪器 135 1095 0799 ...
Страница 58: ...深圳德标仪器 135 1095 0799 ...
Страница 60: ...深圳德标仪器 135 1095 0799 ...
Страница 78: ...深圳德标仪器 135 1095 0799 ...
Страница 286: ...深圳德标仪器 135 1095 0799 ...
Страница 288: ...深圳德标仪器 135 1095 0799 ...
Страница 304: ...深圳德标仪器 135 1095 0799 ...
Страница 632: ...深圳德标仪器 135 1095 0799 ...