Non Signalling: Power Measurements
CMU-K20/K21/K22/K23/K24
1115.6088.12 4.12
E-11
Repetition
The hotkey
Repetition
determines the repetition mode of the measurement
(Single
Shot
or
Continuous
measurement).
Remote control
CONFigure:POWer[:NORMal][:GMSK]:CONTrol:REPetition
<Repetition>
,<StopCond>,<Stepmode>
Stop
Condition
The
Stop Condition
hotkey sets a stop condition for the measurement
(None
or
On
Limit Failure)
.
Remote control
CONFigure:POWer[:NORMal][:GMSK]:CONTrol:REPetition
<Repetition>,
<StopCond>
,<Stepmode>
Display
Mode
The hotkey
Display Mode
determines the display mode of the measurement curve.
Remote control
no display mode set, the four measurement curves are accessible via
FETCh:ARRAy:POWer[:NORMal][:GMSK][:CURRent]?
FETCh:ARRAy:POWer[:NORMal][:GMSK]:MINimum?
FETCh:ARRAy:POWer[:NORMal][:GMSK]:MAXimum?
FETCh:ARRAy:POWer[:NORMal][:GMSK]:AVERage?
etc.
Statistic
Count
The
Statistic Count
hotkey defines the number of bursts per statistic cycle.
Remote control
CONFigure:POWer[:NORMal][:GMSK]:CONTrol
<Mode>,1 ... 1000 | NONE
Trig. Slot
Offset
The
Trig. Slot Offset
hotkey defines a delay time (integer number of GSM
timeslots) between the trigger time and the measured timeslot (see
Fig. 4-3
on p.
4.18
. In the default setting
(Trig. Slot Offset = 0)
the measured timeslot is
determined by the trigger time. By varying the
Trig. Slot Offset,
an unknown GSM
signal can be analyzed timeslot by timeslot at constant trigger settings.
Remote control
CONFigure:RFANalyzer:MCONTrol:TSOFfset 0 to 7
Analyzer
Level
The
Analyzer Level
softkey controls the level in the RF input signal path and
provides the trigger settings for the measurements.
The input level and trigger settings are also provided in the
Analyzer
tab of the
Connection Control
menu. For a detailed description see section
Table-Oriented
Version
on p.
4.73
ff.
RF Max.
Level
The
RF Max. Level
hotkey sets the maximum expected input level in dBm.
Remote control
[SENSE:]LEVel:MAXimum <Level>
RF
Mode
The
RF Mode
hotkey determines how the input level is defined.
Manual
Manual input via
RF Max. Level
hotkey
Auto
Automatic setting according to the average burst power of the
applied signal.
深圳德标仪器
135-1095-0799
Содержание CMU-K20
Страница 4: ...深圳德标仪器 135 1095 0799 ...
Страница 8: ...深圳德标仪器 135 1095 0799 ...
Страница 10: ...深圳德标仪器 135 1095 0799 ...
Страница 12: ...深圳德标仪器 135 1095 0799 ...
Страница 18: ...深圳德标仪器 135 1095 0799 ...
Страница 24: ...深圳德标仪器 135 1095 0799 ...
Страница 26: ...深圳德标仪器 135 1095 0799 ...
Страница 34: ...深圳德标仪器 135 1095 0799 ...
Страница 58: ...深圳德标仪器 135 1095 0799 ...
Страница 60: ...深圳德标仪器 135 1095 0799 ...
Страница 78: ...深圳德标仪器 135 1095 0799 ...
Страница 286: ...深圳德标仪器 135 1095 0799 ...
Страница 288: ...深圳德标仪器 135 1095 0799 ...
Страница 304: ...深圳德标仪器 135 1095 0799 ...
Страница 632: ...深圳德标仪器 135 1095 0799 ...