Common: MODulation[:PERRor]
CMU-K20/21/22/23/24
1115.6088.12 6.68
E-12
Subsystem MODulation...:TIME
The subsystem
MODulation...:TIME
defines the decoding for the
Modulation
measurement. The
subsystem corresponds to the
Decode
hotkey in the graphical measurement menu
Modulation.
CONFigure:MODulation[:PERRor][:GMSK]:TIME:DECode
<Mode>
Decode
<Mode>
Description of parameters
Def. value
Def. unit
FW vers.
STANdard |
GTBits
The standard bit range is decoded
The guard and tail bits are also decoded
GTBits – V2.15
Description of command
This command selects the type of decoding applied for the determination of phase and frequency errors.
Subsystem SUBarrays:MODulation[:PERRor][:GMSK]
The subsystem
SUBarrays:MODulation[:PERRor][:GMSK]
defines the measurement range and the type
of returned values.
CONFigure:SUBarrays:MODulation[:PERRor][:GMSK] <Mode>,<Start>,<Samples>{,<Start>,<Samples>}
Definition of Subarrays
<Mode>
Description of parameters
Def. value
Def. unit
ALL |
ARIThmetical |
MINimum |
MAXimum |
IVAL,
Return all measurement values
Return arithm. mean value in every range
Return minimum value in every range
Return maximum value in every range
Return single interpolated value at <Start>
ALL –
<Start>
Description of parameters
Def. value
Def. unit
0 bit to 146 ¾ bit,
Start time in current range
0
bit
<Samples>
Description of parameters
Def. value
Def. unit
FW vers.
1 to 588
Number of samples in current range
588
–
V2.00
Description of command
This command configures the
READ:SUBarrays..., FETCh:SUBarrays...,
and
SAMPle:SUBarrays:MODulation[:PERRor][:GMSK]
commands. It restricts the measurement to up to 32
subranges where either all measurement results (the number of which is given by the second numerical
parameter) or a single statistical value is returned. The subranges are defined by the start time and the number
of test points which are located on a fixed, equidistant grid with a step width of ¼ bit. If <Start> does not coincide
with a test point then the range will start at the next test point that is larger than <Start>.
For <Mode> = IVAL, the <Samples> parameter is ignored and the CMU returns a single measurement value
corresponding to the abscissa value <Start>. If <Start> is located beween two test points with valid results then
the result is calculated from the results at these two adjacent test points by linear interpolation.
The subranges may overlap but must be within the total range of the
MODulation[:PERRor][:GMSK]
measurement. Test points outside this range are not measured (result
NAN)
and do not enter into the
ARIThmetical, MINimum and MAXimum values.
By default, only one range corresponding to the total measurement range is used and all measurement values
are returned.
深圳德标仪器
135-1095-0799
Содержание CMU-K20
Страница 4: ...深圳德标仪器 135 1095 0799 ...
Страница 8: ...深圳德标仪器 135 1095 0799 ...
Страница 10: ...深圳德标仪器 135 1095 0799 ...
Страница 12: ...深圳德标仪器 135 1095 0799 ...
Страница 18: ...深圳德标仪器 135 1095 0799 ...
Страница 24: ...深圳德标仪器 135 1095 0799 ...
Страница 26: ...深圳德标仪器 135 1095 0799 ...
Страница 34: ...深圳德标仪器 135 1095 0799 ...
Страница 58: ...深圳德标仪器 135 1095 0799 ...
Страница 60: ...深圳德标仪器 135 1095 0799 ...
Страница 78: ...深圳德标仪器 135 1095 0799 ...
Страница 286: ...深圳德标仪器 135 1095 0799 ...
Страница 288: ...深圳德标仪器 135 1095 0799 ...
Страница 304: ...深圳德标仪器 135 1095 0799 ...
Страница 632: ...深圳德标仪器 135 1095 0799 ...