For GSM850, the following four default sequences are available:
Sequence A:
130, 134, 138, 143, 145, 149, 153, 157, 163, 167, 171, 176, 181, 185, 189, 193, 198, 202, 205, 209, 211,
215, 217, 219, 221, 227, 231, 234, 238, 242, 246, 250, OFF, ... OFF
Sequence B:
132, 135, 139, 142, 144, 148, 151, 156, 161, 164, 166, 173, 177, 179, 182, 186, OFF,...OFF
Sequence C:
195, 203, 206, 210, 213, 216, 220, 223, 226, 229, 232, 235, 239, 243, 245, 249, OFF, ... OFF
Sequence D:
128, 190, 251, OFF, ... OFF
For GSM900, the following four default sequences are available:
Sequence A
:
5, 9, 16, 23, 28, 30, 34, 39, 44, 48, 51, 54, 59, 64, 69, 73, 75, 79, 82, 86, 89, 92, 96, 98, 101, 106, 109, 113,
117, 121, OFF, ... OFF
Sequence B:
4, 12, 17, 19, 23, 25, 29, 33, 36, 41, 45, 47, 53, 61, 63, OFF,...OFF
Sequence C
:
65, 68, 72, 76, 79, 81, 84, 88, 91, 93, 98, 102, 105, 112, 118, OFF, ... OFF
Sequence D:
1, 62, 124, OFF, ... OFF
For GSM1800, the following four default sequences are available:
Sequence A
:
533, 559, 568, 592, 604, 617, 631, 642, 678, 697, 722, 743, 759, 796, 811, 824, OFF, ... OFF
Sequence B:
513, 518, 527, 533, 541, 545, 553, 562, 570, 577, 585, 597, OFF,...OFF
Sequence C
:
755, 761, 773, 777, 788, 796, 801, 807, 816, 824, 829, 833, 847 OFF, ... OFF
Sequence D:
512, 698, 885, OFF, ... OFF
For GSM1900, the following four default sequences are available:
Sequence A
:
533, 559, 568, 592, 604, 617, 631, 642, 678, 697, 722, 743, 759, 796, OFF, ... OFF
Sequence B:
513, 518, 527, 533, 541, 545, 553, 562, 570, 577, 585, 597, OFF,...OFF
Sequence C
:
755, 761, 773, 777, 788, 796, 801, 807, OFF, ... OFF
Sequence D:
512, 660, 810, OFF, ... OFF
CONFigure:BSSignal[:CSWitched][:TCH]:MSLot:RLEVel <
Level
>
Reference Level
PROCedure:BSSignal[:CSWitched][:TCH]:MSLot:RLEVel <
Level
>
<Level>
Description of parameters
Def. value
Def. unit
FW vers.
–137 dBm to –27 dBm
–137 dBm to –10 dBm
–90 dBm to +13 dBm
RF1 level in used timeslot
RF2 level in used timeslot
RF3 OUT level in used timeslot
–85
–85
–85
dBm
dBm
dBm
V3.05
Description of command
Sig. State
This command defines the reference value for the individual downlink (BS) signal levels. See
command
PROCedure:SIGNalling[:CSWitched][:TCH]:MSLot:BS:SCONfig:INDividual
on p.
6.181
.
See also note on
Current vs. default values
on p.
6.174
.
all
深圳德标仪器
135-1095-0799
Содержание CMU-K20
Страница 4: ...深圳德标仪器 135 1095 0799 ...
Страница 8: ...深圳德标仪器 135 1095 0799 ...
Страница 10: ...深圳德标仪器 135 1095 0799 ...
Страница 12: ...深圳德标仪器 135 1095 0799 ...
Страница 18: ...深圳德标仪器 135 1095 0799 ...
Страница 24: ...深圳德标仪器 135 1095 0799 ...
Страница 26: ...深圳德标仪器 135 1095 0799 ...
Страница 34: ...深圳德标仪器 135 1095 0799 ...
Страница 58: ...深圳德标仪器 135 1095 0799 ...
Страница 60: ...深圳德标仪器 135 1095 0799 ...
Страница 78: ...深圳德标仪器 135 1095 0799 ...
Страница 286: ...深圳德标仪器 135 1095 0799 ...
Страница 288: ...深圳德标仪器 135 1095 0799 ...
Страница 304: ...深圳德标仪器 135 1095 0799 ...
Страница 632: ...深圳德标仪器 135 1095 0799 ...