Non Signalling: Connection Control
CMU-K20/K21/K22/K23/K24
1115.6088.12 4.74
E-11
The following settings are not provided in the table-oriented version of the
Analyzer
tab:
Default Settings
The
Default All Settings
switch assigns default values to all settings in the
Analyzer
tab (the default values are quoted in the command description in chapter 6 of this
manual).
Remote control
RFANalyzer:DEFault
Analyzer Level –
RF Mode
The
Analyzer Level
table section
sets the maximum input level that can be
measured. The maximum input level is displayed next to the softkey
Max. Level
in
the main menu
Analyzer/Generator
(see page
4.2
) and in the parameter lines
above the graphical measurement menus. Two alternative
RF Modes
for defining
this value are provided:
Manual
Manual input of maximum input level in the
RF Max. Level
field
Auto
Automatic setting of maximum input level
(autoranging)
according to the peak power (PEP) of applied signal
Remote control
[SENSe:]LEVel:MODE <Mode>
Analyzer Level –
RF Max. Level
The maximum expected input level can be entered in the
RF Max. Level
input field.
Input levels exceeding the
RF Max. Level
overdrive the input path and cause
invalid results (“– – –“).
External input
attenuation
The range of
RF Max. Level
values depends on the RF input used. If an external
input attenuation is reported to the instrument to compensate for a known path loss
(see section
AF/RF Connectors (Connection Control – AF/RF)
on page
4.80
), all
levels measured are referenced to the output of the DUT and therefore shifted with
respect to the actual level at the input connectors of the CMU. The level ranges for
the input connectors are shifted as well.
Error messages
If the value determined for
RF Max. Level
is too high or too low, a window with the
error message
"<Max_Level> is out of range. <permissible max. value> is limit
."
and three fields will appear:
Accept
The permissible max. value is accepted as
RF Max. Level
,
Re-edit
RF Max. Level
is entered once again,
Cancel
The last valid input value is maintained.
When switching over to another input, the current value of
RF Max. Level
is
automatically adapted, if required:
•
Towards lower values to the maximum value of the new input,
•
Towards upper values to the minimum value of the new input.
Note:
A maximum input level can be entered even if automatic level setting
(autoranging) is selected. The entered level is used as a start value for
the autoranging routine and is also important to ensure safe
switchover to manual setting.
Remote control
[SENSe:]LEVel:MAXimum <Level>
深圳德标仪器
135-1095-0799
Содержание CMU-K20
Страница 4: ...深圳德标仪器 135 1095 0799 ...
Страница 8: ...深圳德标仪器 135 1095 0799 ...
Страница 10: ...深圳德标仪器 135 1095 0799 ...
Страница 12: ...深圳德标仪器 135 1095 0799 ...
Страница 18: ...深圳德标仪器 135 1095 0799 ...
Страница 24: ...深圳德标仪器 135 1095 0799 ...
Страница 26: ...深圳德标仪器 135 1095 0799 ...
Страница 34: ...深圳德标仪器 135 1095 0799 ...
Страница 58: ...深圳德标仪器 135 1095 0799 ...
Страница 60: ...深圳德标仪器 135 1095 0799 ...
Страница 78: ...深圳德标仪器 135 1095 0799 ...
Страница 286: ...深圳德标仪器 135 1095 0799 ...
Страница 288: ...深圳德标仪器 135 1095 0799 ...
Страница 304: ...深圳德标仪器 135 1095 0799 ...
Страница 632: ...深圳德标仪器 135 1095 0799 ...