Signalling: POWer[:NORMal] CMU-K20/21/22/23/24
1115.6088.12 6.126
E-12
The default settings for 8MSK modulation (EDGE channels, keyword
:EPSK
) are given in the table below:
for Enable
for Table
Start
Stop
Start
Stop
Start
Stop
Suffix Enable
Time Time rel.Level
rel.Level abs.Level abs.Level
Visibility
1
ON
–10.0 symb
½ symb
OFF
OFF
OFF
OFF
OFF
2
ON
½ symb
1 symb
–2.0 dB
–2.0 dB
OFF
OFF
ON
3
ON
1 symb
1 ½ symb
0.0 dB
0.0 dB
OFF
OFF
ON
4
ON
1 ½ symb
146 ½ symb
–15.0 dB
–15.0 dB
OFF
OFF
ON
5
ON
146 ½ symb
147 symb
0.0 dB
0.0 dB
OFF
OFF
ON
6
ON
147 symb
147 ½ symb
–2.0 dB
–2.0 dB
OFF
OFF
ON
7
ON
147 ½ symb
156 ¾ symb
OFF
OFF
OFF
OFF
OFF
8
OFF
OFF OFF OFF
OFF OFF OFF OFF
...
16
OFF
OFF OFF OFF
OFF OFF OFF OFF
CONFigure:POWer[:NORMal][:GMSK]:LIMit:LINE:UPPer<AreaNr>:DYNamic<RangeNr>:ENABle
CONFigure:POWer[:NORMal]:EPSK:LIMit:LINE:UPPer<AreaNr>:DYNamic<RangeNr>:ENABle
<Enable>
CONFigure:POWer[:NORMal][:GMSK]:LIMit:LINE:UPPer<AreaNr>:DYNamic<RangeNr>
CONFigure:POWer[:NORMal]:EPSK:LIMit:LINE:UPPer<AreaNr>:DYNamic<RangeNr>
<fromTPCL>,
<toTPCL>, <Correction>, <Enable>
Dynamic Correction
<fromPCL>
Description of parameters
Def. value
Def. unit
0 to 31 |
MAX | OFF
First PCL for which area <AreaNr> is changed
See table below
PCL
<toPCL>
Description of parameters
Def. value
Def. unit
0 to 31 | MAX
| OFF
Last PCL for which area <AreaNr> is changed.
See table below
PCL
<Correction>
Description of parameters
Def. value
Def. unit
-10 dB to
+10 dB | OFF
Correction value (relative) for the limit line in area
<AreaNr> applied for all PCLs between
<fromPCL> and <toPCL> (including)
See table below
dB
<Enable>
Description of parameters
Def. value
Def. unit
FW vers.
ON | OFF
Enable or disable dynamic correction in the
current limit line area and PCL range
See table below
–
V1.15
Description of command
These command activates and defines dynamic correction of the upper limit line of area <AreaNr> (<AreaNr> = 1
to 16) depending on the PCL range <RangeNr> (<RangeNr> = 1 to 10 for each area).
MAX
denotes the
maximum output power (smallest PCL) of the mobile phone under test.
In the areas no. 3 and 6, the following ranges are defined:
Range
from TPCL
to TPCL
Correction
Enable
1 16
16
+2.0
dB
ON
2 17
17
+4.0
dB
ON
3 18
19
+5.0
dB
ON
4 OFF
OFF
0.0
dB
OFF
5 OFF
OFF
0.0
dB
OFF
6 OFF
OFF
0.0
dB
OFF
7 OFF
OFF
0.0
dB
OFF
8 OFF
OFF
0.0
dB
OFF
9 OFF
OFF
0.0
dB
OFF
10 OFF
OFF
0.0
dB
OFF
In the remaining areas, the dynamic limit line correction is disabled in all ranges.
深圳德标仪器
135-1095-0799
Содержание CMU-K20
Страница 4: ...深圳德标仪器 135 1095 0799 ...
Страница 8: ...深圳德标仪器 135 1095 0799 ...
Страница 10: ...深圳德标仪器 135 1095 0799 ...
Страница 12: ...深圳德标仪器 135 1095 0799 ...
Страница 18: ...深圳德标仪器 135 1095 0799 ...
Страница 24: ...深圳德标仪器 135 1095 0799 ...
Страница 26: ...深圳德标仪器 135 1095 0799 ...
Страница 34: ...深圳德标仪器 135 1095 0799 ...
Страница 58: ...深圳德标仪器 135 1095 0799 ...
Страница 60: ...深圳德标仪器 135 1095 0799 ...
Страница 78: ...深圳德标仪器 135 1095 0799 ...
Страница 286: ...深圳德标仪器 135 1095 0799 ...
Страница 288: ...深圳德标仪器 135 1095 0799 ...
Страница 304: ...深圳德标仪器 135 1095 0799 ...
Страница 632: ...深圳德标仪器 135 1095 0799 ...