Common: POWer:FRAMe
CMU-K20/21/22/23/24
1115.6088.12 6.52
E-12
Subsystem SUBarrays:POWer:FRAMe
The subsystem
SUBarrays:POWer:FRAMe
defines the measurement range and the type of return
values.
CONFigure:SUBarrays:POWer:FRAMe <Mode>,<Start>,<Samples>{,<Start>,<Samples>}
Definition of Subarrays
<Mode>
Description of parameters
Def. value
Def. unit
ALL |
ARIThmetical |
MINimum |
MAXimum |
IVAL,
Return all measurement values
Return arithm. mean value in every range
Return minimum value in every range
Return maximum value in every range
Return single interpolated value at <Start>
ALL –
<Start>
Description of parameters
Def. value
Def. unit
0 to 255,
Start time in current range
0
–
<Samples>
Description of parameters
Def. value
Def. unit
FW vers.
1 to 256
Number of samples in current range
128
–
V2.15
Description of command
This command configures the
READ:SUBarrays..., FETCh:SUBarrays...,,
and
SAMPle:SUBarrays:POWer:FRAMe
commands. It restricts the measurement to up to 32 subranges where
either all measurement results (the number of which is given by the second numerical parameter) or a single
statistical value is returned. The subranges are defined by the current number of the first slot and the number of
slots within a subrange. If <Start> does not coincide with a test point then the range will start at the next test point
that is larger than <Start>.
For <Mode> = IVAL, the <Samples> parameter is ignored and the CMU returns a single measurement value
corresponding to the abscissa value <Start>. If <Start> is located beween two test points with valid results then
the result is calculated from the results at these two adjacent test points by linear interpolation.
The subranges may overlap but must be within the total range of the
P/Frame
measurement. Test points outside
this range are not measured (result
NAN)
and do not enter into the ARIThmetical, MINimum and MAXimum
values.
By default, only one range corresponding to the total measurement range is used and all measurement values
are returned.
Subsystem POWer:FRAMe
The subsystem
POWer:FRAMe
measures the power versus frame and returns the results. The
subsystem corresponds to the measurement menu
P/Frame.
深圳德标仪器
135-1095-0799
Содержание CMU-K20
Страница 4: ...深圳德标仪器 135 1095 0799 ...
Страница 8: ...深圳德标仪器 135 1095 0799 ...
Страница 10: ...深圳德标仪器 135 1095 0799 ...
Страница 12: ...深圳德标仪器 135 1095 0799 ...
Страница 18: ...深圳德标仪器 135 1095 0799 ...
Страница 24: ...深圳德标仪器 135 1095 0799 ...
Страница 26: ...深圳德标仪器 135 1095 0799 ...
Страница 34: ...深圳德标仪器 135 1095 0799 ...
Страница 58: ...深圳德标仪器 135 1095 0799 ...
Страница 60: ...深圳德标仪器 135 1095 0799 ...
Страница 78: ...深圳德标仪器 135 1095 0799 ...
Страница 286: ...深圳德标仪器 135 1095 0799 ...
Страница 288: ...深圳德标仪器 135 1095 0799 ...
Страница 304: ...深圳德标仪器 135 1095 0799 ...
Страница 632: ...深圳德标仪器 135 1095 0799 ...