Common: POWer:MSLot
CMU-K20/21/22/23/24
1115.6088.12 6.62
E-12
Tolerance Template
The subsystem
POWer:MSLot:AREA:LIMit...
contains the commands to return the current position of
the multislot tolerance template and the curve. The subsystem has no equivalent in manual control,
however, the current template is indicated in the graphical
P/t Multislot
digaram
.
[SENSe:]ARRay:POWer:MSLot:AREA:LIMit:UPPer:TIME?
Time of all Areas
[SENSe:]ARRay:POWer:MSLot:AREA:LIMit:LOWer:TIME?
Returned values
Value range
Def. value
Def. unit
FW vers.
–180 symb. to +520 symb. | OFF,
–180 symb. to +520 symb. | OFF,
...
–180 symb. to +520 symb. | OFF,
–180 symb. to +520 symb. | OFF
Start time in area no. 1
Stop time in area no. 1
Start time in area no. n
Stop time in area no. n
NAN
NAN
NAN
NAN
symbols
V3.10
Description of command
These commands return the time of all areas of the multislot tolerance template, relative to the start of the
measured timeslot
(Meas. Slot)
. OFF means that the limit line and limit check in an area is switched off. The
number of areas and thus the number of output values varies with the number of measured slots and the
definition of the single slot template. The maximum allowed number of output values is 2 x 64.
[SENSe:]ARRay:POWer:MSLot:AREA:LIMit:UPPer:LEVel?
Level of all Areas
[SENSe:]ARRay:POWer:MSLot:AREA:LIMit:LOWer:LEVel?
Returned values
Value range
Def. value
Def. unit
FW vers.
–100.0 dB to + 20.0 dB | OFF,
–100.0 dB to + 20.0 dB | OFF,
...
–100.0 dB to + 20.0 dB | OFF,
–100.0 dB to + 20.0 dB | OFF
Start level in area no. 1
Stop level in area no. 1
Start level in area no. n
Stop level in area no. n
NAN
NAN
NAN
NAN
dB
dB
dB
dB
V3.10
Description of command
These commands return the level of all areas of the multislot tolerance template, relative to the useful level of the
measured timeslot
(Meas. Slot)
. OFF means that the limit line and limit check in an area is switched off. The
number of areas and thus the number of output values varies with the number of measured slots and the
definition of the single slot template. The maximum allowed number of output values is 2 x 64.
[SENSe:]ARRay:POWer:MSLot:AREA:LIMit:UPPer:INFO?
Timeslot of all Areas
[SENSe:]ARRay:POWer:MSLot:AREA:LIMit:LOWer:INFO?
Returned values
Value range
Def. value
Def. unit
FW vers.
–1 | 0 | 1 | 2,
...
–1 | 0 | 1 | 2*
)
Timeslot of area no. 1
Timeslot of area no. n
NAN
NAN
dB
dB
V3.10
Description of command
These commands return the timeslot of all areas of the multislot tolerance template, relative to the measured
timeslot
(Meas. Slot,
slot no. 0
)
. The number of areas and thus the number of output values varies with the
number of measured slots and the definition of the single slot template. The maximum allowed number of output
values is 64.
*) If the timeslots no. –2 or +3 are active and if an area overlaps to one of these slots, the query may also return –2 or +3.
深圳德标仪器
135-1095-0799
Содержание CMU-K20
Страница 4: ...深圳德标仪器 135 1095 0799 ...
Страница 8: ...深圳德标仪器 135 1095 0799 ...
Страница 10: ...深圳德标仪器 135 1095 0799 ...
Страница 12: ...深圳德标仪器 135 1095 0799 ...
Страница 18: ...深圳德标仪器 135 1095 0799 ...
Страница 24: ...深圳德标仪器 135 1095 0799 ...
Страница 26: ...深圳德标仪器 135 1095 0799 ...
Страница 34: ...深圳德标仪器 135 1095 0799 ...
Страница 58: ...深圳德标仪器 135 1095 0799 ...
Страница 60: ...深圳德标仪器 135 1095 0799 ...
Страница 78: ...深圳德标仪器 135 1095 0799 ...
Страница 286: ...深圳德标仪器 135 1095 0799 ...
Страница 288: ...深圳德标仪器 135 1095 0799 ...
Страница 304: ...深圳德标仪器 135 1095 0799 ...
Страница 632: ...深圳德标仪器 135 1095 0799 ...