Signalling: Receiver Quality Measurements
CMU-K20/K21/K22/K23/K24
1115.6088.12 4.128
E-12
Dual-Limit Test
The single-limit BER test described above can be extended to restrict the BER to a
band between two limits. For practical reasons, the lower band limit is calculated as
<BER Limit> (1 – n %), the upper band limit is calculated as <BER Limit> (1 + n %).
Factors of 10 %, 20 % and 30 % are provided; see
Result Window
parameter.
Fig. 4-59 Dual-limit confidence BER
In a dual-limit test, the DUT is failed if it is either too good or too bad. For more
information refer to the test specification.