Common: NPOWer
CMU-K20/21/22/23/24
1115.6088.12 6.40
E-12
CONFigure:NPOWer:CONTrol:REPetition <
Repetition
>
,<StopCond>,<Stepmode>
Test cycles
<Repetition>
Description of parameters
Def. value
Def. unit
CONTinuous |
SINGleshot |
1 ... 10000
Continuous measurement (until
STOP
or
ABORT)
Single shot measurement (until
Status = RDY
)
Multiple measurement
(
counting
, until
Status = STEP | RDY)
SING
–
<StopCond>
Description of parameters
Def. value
Def. unit
SONerror |
NONE
Start measurement in case of error
(stop on error)
Continue measurement even in case of error
NONE –
<Stepmode>
Description of parameters
Def. value
Def. unit
FW vers.
STEP |
NONE
Interrupt measurement after each statistics cycle
Continue measurement according to its rep. mode
NONE –
V3.05
Description of command
This command determines the number of statistics cycles, the stop condition and the stepping mode for the
measurement.
Note:
In the case of READ commands (
READ:
...), the <Repetition> parameter has no effect; the
measurement is always stopped after a single shot.
Measured Values – Subsystem NPOWer?
The subsystem
NPOWer?
retrieves the results of the narrow-band power measurement.
READ[:SCALar]:NPOWer?
Start single shot measurement and return results
FETCh[:SCALar]:NPOWer?
Read out measurement results (unsynchronized)
SAMPle[:SCALar]:NPOWer?
Read out measurement results (synchronized)
Returned values
Value range
Def. value
Def. unit
FW vers.
Avg. Power of Current evaluation period,
Min. Power of Current evaluation period,
Max. Power of Current evaluation period,
Avg. Power ref. to the last stat. cycle,
Min. Power of the entire measurement,
Max. Power of the entire measurement
–137 dBm to +53 dBm
–137 dBm to +53 dBm
–137 dBm to +53 dBm
–137 dBm to +53 dBm
–137 dBm to +53 dBm
–137 dBm to +53 dBm
NAN
NAN
NAN
NAN
NAN
NAN
dBm
dBm
dBm
dBm
dBm
dBm
V3.05
Description of command
These commands are always queries. They start the
NPOWer
measurement and return the results.
深圳德标仪器
135-1095-0799
Содержание CMU-K20
Страница 4: ...深圳德标仪器 135 1095 0799 ...
Страница 8: ...深圳德标仪器 135 1095 0799 ...
Страница 10: ...深圳德标仪器 135 1095 0799 ...
Страница 12: ...深圳德标仪器 135 1095 0799 ...
Страница 18: ...深圳德标仪器 135 1095 0799 ...
Страница 24: ...深圳德标仪器 135 1095 0799 ...
Страница 26: ...深圳德标仪器 135 1095 0799 ...
Страница 34: ...深圳德标仪器 135 1095 0799 ...
Страница 58: ...深圳德标仪器 135 1095 0799 ...
Страница 60: ...深圳德标仪器 135 1095 0799 ...
Страница 78: ...深圳德标仪器 135 1095 0799 ...
Страница 286: ...深圳德标仪器 135 1095 0799 ...
Страница 288: ...深圳德标仪器 135 1095 0799 ...
Страница 304: ...深圳德标仪器 135 1095 0799 ...
Страница 632: ...深圳德标仪器 135 1095 0799 ...