CMU-K20/K21/K22/K23/K24
Non Signalling: Power Measurements
1115.6088.12 4.15
E-11
Remote control
CONFigure:RFGenerator:MODulation:TSEQuence:SELection <TSC>
Bit
Modulation
The
Bit Modulation
hotkey selects a bit sequence to be modulated onto the
generator signal.
Remote control
CONFigure:RFGenerator:MODulation:BIT:SELection <Sequence>
Trans-
mission
The
Transmission
hotkey determines the shape of the generator signal (burst
signal or continuous wave with constant level). An 8PSK-modulated signal is
always bursted.
Remote control
CONFigure:RFGenerator:MODulation:TRANsmission <Mode>
Marker
Display
The
Marker/Display
softkey positions up to 3 markers and a D-line in the test
diagram and displays their values.
If pressed once again, the selected
Marker/Display
softkey changes to the
Display/Marker
softkey, see below.
Markers
are graphical tools for marking points on the measurement curve and
for numerical output of measured values. The measurement menu
Power provides a reference marker and two further markers which
permit to measure differences (delta marker 1 and 2).
The coordinates of the three markers are indicated in the format
Ordinate value (level)/abscissa value (time) in a parameter line above
the test diagram. The position of the reference marker is expressed in
absolute units (level in dBm and time in symbol periods), the delta
marker by absolute or relative values (relative level in dB or time
differences from the reference marker).
D-line
The D-line (display line) is a horizontal line that can be positioned on
the test diagram at will to mark and read out level values.
Ref
The hotkey
Ref. R
switches the reference marker on or off (use the
ON/OFF
key).
The reference marker is represented by the symbol in the test diagram. The
marker position (abscissa) is defined in the input field
Ref. Marker R
. The marker
can be positioned to arbitrary time values. It is switched off in the default setting
(Off).
The marker level is given by the measurement curve at the marker position.
The position of all markers can be varied using the rollkey.
Remote control
No command, screen configuration only.
Rel
The
Rel. 1
hotkey switches the delta marker 1 on or off (use the
ON/OFF
key).
The delta marker 1 is represented by the symbol in the test diagram. The
marker position (abscissa) is defined in the input field
Rel. Marker 1
. The marker
can be positioned to arbitrary time values. If its position is outside the diagram area
it will be invisible and its coordinates will be “<abscissa_value> / - - -“. The marker
is switched off in the default setting
(Off).
The marker level is given by the
measurement curve at the marker position.
The toggle switch
Rel 1 Config
pops up when the hotkey is pressed for the second
time. It defines whether the position of delta marker 1 is measured and indicated in
absolute units (dBm) or relative to the reference marker.
深圳德标仪器
135-1095-0799
Содержание CMU-K20
Страница 4: ...深圳德标仪器 135 1095 0799 ...
Страница 8: ...深圳德标仪器 135 1095 0799 ...
Страница 10: ...深圳德标仪器 135 1095 0799 ...
Страница 12: ...深圳德标仪器 135 1095 0799 ...
Страница 18: ...深圳德标仪器 135 1095 0799 ...
Страница 24: ...深圳德标仪器 135 1095 0799 ...
Страница 26: ...深圳德标仪器 135 1095 0799 ...
Страница 34: ...深圳德标仪器 135 1095 0799 ...
Страница 58: ...深圳德标仪器 135 1095 0799 ...
Страница 60: ...深圳德标仪器 135 1095 0799 ...
Страница 78: ...深圳德标仪器 135 1095 0799 ...
Страница 286: ...深圳德标仪器 135 1095 0799 ...
Страница 288: ...深圳德标仪器 135 1095 0799 ...
Страница 304: ...深圳德标仪器 135 1095 0799 ...
Страница 632: ...深圳德标仪器 135 1095 0799 ...