Measurement Groups (Non Signalling only) CMU-K20/K21/K22/K23/K24
1115.6088.12 6.28
E-12
Subsystem SUBarrays:POWer:MPR
The subsystem
SUBarrays:POWer:MPR
defines the measurement range and the type of output values.
CONFigure:SUBarrays:POWer[:NORMal][:GMSK]:MPR <Mode>,<Start>,<Samples>{,<Start>,<Samples>}
Definition of Subarrays
<Mode>
Description of parameters
Def. value
Def. unit
ALL |
ARIThmetical |
MINimum |
MAXimum |
IVAL,
Return all measurement values
Return arithm. mean value in every range
Return minimum value in every range
Return maximum value in every range
Return single interpolated value at <Start>
ALL –
<Start>
Description of parameters
Def. value
Def. unit
–10 bit to 156 ¾ bit,
Start time in current range
–10
bit
<Samples>
Description of parameters
Def. value
Def. unit
FW vers.
1 to 668
Number of samples in current range
668
–
V2.00
Description of command
This command configures the
READ:SUBarrays..., FETCh:SUBarrays...,,
and
SAM-
Ple:SUBarrays:POWer:MPR
commands. It restricts the measurement to up to 32 subranges where either all
measurement results (the number of which is given by the second numerical parameter) or a single statistical
value is returned. The subranges are defined by the start time and the number of test points which are located on
a fixed, equidistant grid with a step width of ¼ bit. If <Start> does not coincide with a test point then the range will
start at the next test point that is larger than <Start>.
For <Mode> = IVAL, the <Samples> parameter is ignored and the CMU returns a single measurement value
corresponding to the abscissa value <Start>. If <Start> is located beween two test points with valid results then
the result is calculated from the results at these two adjacent test points by linear interpolation.
The subranges may overlap but must be within the total range of the
Power
measurement. Test points outside
this range are not measured (result
NAN)
and do not enter into the ARIThmetical, MINimum and MAXimum val-
ues.
By default, only one range corresponding to the total measurement range is used and all measurement values
are returned.
深圳德标仪器
135-1095-0799
Содержание CMU-K20
Страница 4: ...深圳德标仪器 135 1095 0799 ...
Страница 8: ...深圳德标仪器 135 1095 0799 ...
Страница 10: ...深圳德标仪器 135 1095 0799 ...
Страница 12: ...深圳德标仪器 135 1095 0799 ...
Страница 18: ...深圳德标仪器 135 1095 0799 ...
Страница 24: ...深圳德标仪器 135 1095 0799 ...
Страница 26: ...深圳德标仪器 135 1095 0799 ...
Страница 34: ...深圳德标仪器 135 1095 0799 ...
Страница 58: ...深圳德标仪器 135 1095 0799 ...
Страница 60: ...深圳德标仪器 135 1095 0799 ...
Страница 78: ...深圳德标仪器 135 1095 0799 ...
Страница 286: ...深圳德标仪器 135 1095 0799 ...
Страница 288: ...深圳德标仪器 135 1095 0799 ...
Страница 304: ...深圳德标仪器 135 1095 0799 ...
Страница 632: ...深圳德标仪器 135 1095 0799 ...