Signalling: Receiver Quality
CMU-K20/21/22/23/24
1115.6088.12 6.170
E-12
Receiver Quality – Block Error Ratio
The subsystem
RXQuality:BLER
contains the commands for the Block Error Ratio (BLER)
measurement. The subsystem corresponds to the main menu
Receiver Quality,
application
BLER
and
the corresponding parts of the associated popup menu
Receiver Quality Configuration.
Control of Measurement – Subsystem RXQuality:BLER
The subsystem
RXQuality:BLER
controls the BLER measurement.
INITiate:RXQuality:BLER
Start new measurement
⇒
RUN
ABORt:RXQuality:BLER
Abort running measurement and switch off
⇒
OFF
STOP:RXQuality:BLER
Stop measurement
⇒
STOP
CONTinue:RXQuality:BLER
Next measurement step (only
stepping mode
)
⇒
RUN
Description of command
FW vers.
These commands have no query form. They start or stop the BLER measurement, setting it to the
status indicated in the top right column.
V3.10
CONFigure:RXQuality:BLER:EREPorting <
Mode
>
Event Reporting
<Mode>
Description of parameters
Def. value
Def. unit
FW vers.
SRQ |
SOPC |
SRSQ |
OFF
Service request
Single operation complete
SRQ and SRSQ
No reporting
OFF –
V3.10
Description of command
This command defines the events generated when the measurement is terminated or stopped
(event reporting,
see chapter 5).
FETCh:RXQuality:BLER:STATus?
Measurement Status
Return
Description of parameters
Def. value
Def. unit
FW vers.
OFF |
RUN |
STOP |
ERR |
STEP |
RDY,
1 to 2000 |
NONE
Measurement in the
OFF
state (
*RST
or
ABORt
)
Running (after
INITiate, CONTinue
or
READ
)
Stopped (
STOP
)
OFF
(could not be started)
Stepping mode
(<stepmode>=STEP)
Stopped according to repetition mode and stop cond.
Counter for current evaluation period (RLC blocks)
Statistic count set to OFF (only one block)
OFF
NONE
–
–
V3.10
Description of command
This command is always a query. It returns the status of the measurement (see chapter 5).
深圳德标仪器
135-1095-0799
Содержание CMU-K20
Страница 4: ...深圳德标仪器 135 1095 0799 ...
Страница 8: ...深圳德标仪器 135 1095 0799 ...
Страница 10: ...深圳德标仪器 135 1095 0799 ...
Страница 12: ...深圳德标仪器 135 1095 0799 ...
Страница 18: ...深圳德标仪器 135 1095 0799 ...
Страница 24: ...深圳德标仪器 135 1095 0799 ...
Страница 26: ...深圳德标仪器 135 1095 0799 ...
Страница 34: ...深圳德标仪器 135 1095 0799 ...
Страница 58: ...深圳德标仪器 135 1095 0799 ...
Страница 60: ...深圳德标仪器 135 1095 0799 ...
Страница 78: ...深圳德标仪器 135 1095 0799 ...
Страница 286: ...深圳德标仪器 135 1095 0799 ...
Страница 288: ...深圳德标仪器 135 1095 0799 ...
Страница 304: ...深圳德标仪器 135 1095 0799 ...
Страница 632: ...深圳德标仪器 135 1095 0799 ...